BioTek Synergy MX Multi-Mode Microplate Reader
Detection Modes Absorbance, Fluorescence Intensity, Luminescence
Microplate Formats 1- to 384- well
Read Types Endpoint, Kinetic, Spectral Scanning, Well-Area Scanning
- Fully monochromator-based with quadruple grating architecture for ultra fine-tuned performance
- Compatible with Take3 Multi-Volume Plate with 2 µL microspots for low volume assays
- High performance fluorescence (top read only), absorbance and dedicated luminescence detection system accommodates all life-science research microplate applications
- Automated Z-height adjustment accommodates variable microplate heights and liquid levels with 100 μm resolution
- Variable bandpass selection system adds flexibility
- Cuvette port to measure standard 1-cm path cuvettes
- Controlled by the powerful BioTek Gen5 software, facilitating easy data collection, analysis, exporting, and reporting
The BioTek Synergy MX Multi-Mode Microplate Reader is an advanced, versatile laboratory instrument designed for various life-science research applications. This state-of-the-art device features a fully monochromator-based system with quadruple grating architecture, ensuring ultra-fine-tuned performance across a range of functions. It is compatible with the Take3 Multi-Volume Plate, which includes 2 µL microspots for conducting low volume assays efficiently. Key features of the Synergy MX include high-performance fluorescence detection (top read only), absorbance, and a dedicated luminescence detection system. Its automated Z-height adjustment can accommodate varying microplate heights and liquid levels with a precise 100 μm resolution. The variable bandpass selection system adds further flexibility to its operations. Additionally, the device includes a cuvette port for standard 1-cm path cuvettes.
The absorbance specifications of the Synergy MX are impressive. It utilizes a Xenon flash lamp as its light source and a double monochromator for wavelength selection, covering a range from 230 to 999 nm in 1 nm increments. The device ensures photometric accuracy of < 1% at 2.0 OD and repeatability of < 0.5% at 2.0 OD, with a dynamic range from 0 to 4.0 OD. For fluorescence intensity measurements, the device also uses a Xenon flash lamp and a dedicated red-shifted PMT detector, with a wavelength range of 250 to 900 nm and a remarkable sensitivity level. The luminescence capabilities of the Synergy MX are equally robust, featuring a low noise PMT detection system and a dynamic range of < 6 decades, with exceptional sensitivity for detecting as low as 10 amol of ATP.
This microplate reader can accommodate formats from 1 to 384-well plates and PCR plates, and offers various read methods such as endpoint, kinetic, spectral scanning, and well-area scanning. It also includes plate shaking features and precise temperature control from Ambient +4°C to 65°C. Controlled by BioTek's powerful Gen5 software, the Synergy MX facilitates easy data collection, analysis, exporting, and reporting, making it an invaluable tool for modern laboratories engaged in life-science research. Please Note: This is the SMATCL model of the Synergy MX, which DOES NOT feature bottom fluorescence reading or a dual reagent dispenser.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioTek Synergy MX Multi-Mode Microplate Reader includes:
- > Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- > ABS, FLD, LUM, and Self Test Calibration Report
- > Interface Cable
- > Operator's Manual
- > Full 180 Day Part and Labor Warranty
BioTek Synergy MX Multi-Mode Microplate Reader Specifications
|
Absorbance Specifications
|
|
Light Source
|
Xenon flash lamp
|
|
Wavelength Selection
|
Double monochromator
|
|
Wavelength Range
|
230 to 999 nm, in 1 nm increments
|
|
Bandpass
|
2 nm (230-285 nm), 4 nm (>285 nm)
|
|
Pathlength Correction
|
Yes
|
|
Monochromator Wavelength Accuracy
|
± 2 nm
|
|
Monochromator Wavelength Repeatability
|
± 0.2 nm
|
|
Dynamic Range
|
0 to 4.0 OD
|
|
Photometric Resolution
|
0.0001 OD
|
|
Photometric Accuracy
|
< 1% at 2.0 OD typical
|
|
Photometric Linearity
|
< 1% from 0 to 3.0 OD
|
|
Photometric Repeatability
|
< 0.5% at 2.0 OD
|
|
Reading Speed
|
96 wells: 11 seconds
384 wells: 22 seconds
|
|
Fluorescence Intensity Specifications
|
|
Light Source
|
Xenon flash lamp
|
|
Detector
|
Dedicated red shifted PMT
|
|
Wavelength Selection
|
Double grating monochromators (top read only)
|
|
Wavelength Range
|
250 to 900 nm
|
|
Dynamic Range
|
5 decades
|
|
Sensitivity (Top Read Only)
|
Fluorescein 2 pM typical (0.2 fmol/well, 384-well plate)
|
|
Luminescence Specifications
|
|
Dynamic Range
|
< 6 decades
|
|
Sensitivity
|
10 amol ATP typical (flash)
|
|
Detection System
|
Low noise PMT
|
|
Wavelength Range
|
300 to 700 nm
|
|
General Specifications
|
|
Microplate Formats
|
1- to 384- well plates and PCR plates
|
|
Read Methods
|
End point, kinetic, spectral scanning, well-area scanning
|
|
Plate Shaking
|
Yes
|
|
Temperature Control
|
Ambient +4°C to 65°C, ± 0.5°C at 37°C
|
|
Dimensions (W x D x H)
|
17 x 20.9 x 15 in. (43.5 x 53.1 x 38.1 cm)
|
|
Weight
|
70 lb. (31.5 kg.)
|
|
Electrical Requirements
|
100-240V, 50/60Hz
|
|
Computer Specifications
|
|
Processor
|
Intel Core I5-8350U CPU
|
|
RAM
|
16.0 GB
|
|
Hard Drive
|
512GB Internal SSD
|
|
Operating System
|
Windows 11 Professional, 64-bit
|
Frequently Purchased With
- Verify that the PCB and power supply output proper voltages to all internal components.
- Examine the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal readability.
- Inspect all buttons for responsiveness, confirming full functionality and consistency.
- Assess the plate drawer for proper clearance and smooth operation to ensure accurate plate handling.
- Confirm seamless communication between the plate reader and computer software for data integrity.
- Inspect internal wiring and electrical connections for wear or potential faults, addressing any issues found.
- Examine the lamp, replacing it as needed to maintain proper illumination and performance.
- Execute the required validation protocol to confirm system accuracy and compliance.
- Apply all current updates to the data station and operating system to ensure compatibility and stability.
- Conduct a final cosmetic inspection to verify that all external surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
| Task |
Completed? |
Remarks? |
| Verify PCB and power supply output proper voltages | Completed | |
| Check display for brightness, replace dim or dead pixels | Completed | |
| Inspect buttons for full functionality and responsiveness | Completed | |
| Assess plate drawer clearance for accurate plate handling | Completed | |
| Confirm communication between plate reader and software | Completed | |
| Inspect internal wiring and electrical connections for wear | Completed | |
| Examine lamp and replace as necessary for illumination | Completed | |
| Execute validation protocol for system accuracy | Completed | |
| Apply all current updates to data station and OS | Completed | |
| Conduct cosmetic inspection to meet appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|