VWR 370 Hot Plate Stirrer
Temperature Range Ambient to 400C
Speed Range 60 to 2,000 RPM
Platform Size 8 x 8 in (20.3 x 20.3 cm)microplates
- Wide temperature range from ambient to 400°C for diverse heating applications
- Variable stirring speeds from 60 to 2,000 RPM for precise sample mixing
- Compact and robust design suitable for various laboratory environments
- Reliable construction ensures durability and long-term use in demanding conditions
The VWR Hot Plate Stirrer 370 has an 8" x 8" ceramic top plate, heats up to 400C and stirs between 60-2000rpms. All of our equipment is fully tested before being shipped.
The VWR 370 Hot Plate Stirrer includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
VWR 370 Hot Plate Stirrer Specifications
Temperature Range
|
Ambient to 400C
|
Stirrer Speed
|
60 to 2,000 RPM
|
Platform Size
|
8 x 8 in (20.3 x 20.3 cm)
|
Electrical Requirements
|
115V, 50/60Hz, 800W
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for clarity and replace if needed | Completed | |
Examine external regulators for integrity and damage | Completed | |
Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
Check internal wiring for frays, tears, or breakages | Completed | |
Confirm software settings match internal component specs | Completed | |
Inspect and clean fan, replacing if wear is detected | Completed | |
Verify condition of wheels and leveling mounts | Completed | |
Inspect door gaskets and latches for sealing integrity | Completed | |
Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
Confirm 24-hour temperature and CO₂ stability | Completed | |
Conduct cosmetic inspection to ensure appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|