New Brunswick I2500 Incubator Shaker
Temperature Range
Ambient +7°C to 60°C
Platform Dimensions
30 x 18 in
- Large-capacity design accommodates up to 6-liter flasks for high-volume applications
- Advanced microprocessor control ensures precise regulation of speed, temperature, and runtime
- Easy-open top-loading lid allows for quick and convenient sample access and retrieval
- Powerful fan and heating system enable rapid heat-up and consistent temperature recovery
- Durable, high-quality construction ensures reliable long-term laboratory performance
- Bright, clear digital display provides easy monitoring of operational parameters at a glance
- Even airflow distribution maintains uniform conditions for optimal sample processing
- Low-vibration operation enhances stability and minimizes disruptions to sensitive cultures
- Adjustable settings offer flexibility to accommodate a wide range of laboratory applications
The New Brunswick I2500 Incubator Shaker is a
large-capacity environmental shaker that allows for uniform heating and motion
and accurate regulation of shaking speed. Ideal for large-load requirements, the I2500 can accept up to 6 liter flasks and features an easy-open top-loading lid. Microprocessor control precisely regulates temperature, speed, and run time. Temperatures from 7°C above ambient to 60°C are supported by powerful fans and a heater to provide rapid heat up and temperature recovery after door opening.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The New Brunswick I2500 Incubator Shaker includes:
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > 30 x 18" Platform
- > Full 180 Day Parts and Labor Warranty
New Brunswick I2500 Incubator Shaker Specifications
Temperature Range
|
Ambient +7°C to 60°C
|
Temperature Uniformity
|
± 0.25°C @ 37°C
|
Speed Range
|
25 to 500 RPM
|
Orbit
|
1 inch
|
Platform Size
|
30 x 18 in
|
Dimensions (W x D x H)
|
45 x 28 x 37 in
|
Electrical Requirements
|
120 V, 60 Hz
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|