Thermo Scientific Multifuge 1S Centrifuge
- QUIKset and EASYset Speed Control for precise adjustments in 100 or 10 RPM increments
- Run Timer allows operation from 0 to 9 hours and 59 minutes or continuous hold
- RCF Preselection and Quick-Run Functions for easy operation and flexibility
- Automatic Rotor Recognition ensures safe and optimized performance
- Electronic Imbalance Detection adjusts speed based on rotor conditions
- Lid Lock and Interlock Safety System for secure operation
- Non-Refrigerated Model suitable for general lab use across multiple applications
The Thermo Scientific Multifuge 1S Centrifuge is a 1.6 liter capacity non-refrigerated centrifuge with a wide assortment of rotors available. This centrifuge reaches speeds up to 15,000 RPM with maximum RCF of 24,652 x g. The Multifuge 1S allows for easy high volume processing with automatic rotor recognition and quiet operation. With higher RCFs and great flexibility, the Multifuge 1S is ideal for microbiology, pharmaceutical, clinical testing, biotechnology, and general lab use. This rotor reaches a maximum speed of 10,250 RPM.All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Thermo Scientific Multifuge 1S Centrifuge includes:
- > 75002006 Rotor
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Multifuge 1S Centrifuge Specifications
Maximum Speed
|
15,000 RPM
|
Minimum Speed
|
300 RPM
|
Speed Control
|
100 RPM increments in QUIKset
10 RPM increments in EASYset
|
Maximum RCF
|
24,652 x g
|
Maximum Kinetic Entergy
|
34.3 kNm
|
Run Time
|
0 to 9 hr 59 min, hold
|
Functions
|
RCF-preselection, Quick-run, automatic rotor recognition and
imbalance detection
|
Imbalance Recognition
|
Electronic, speed and rotor dependent
|
Safety
|
Lid lock and lid interlock (automatic lid locks)
|
Dimensions (H x W x D)
|
14.4 x 19.7 x 24.2 in (365 x 500 x 615 mm)
|
Electrical Requirements
|
120 VAC, 60 Hz, 12 A
|
Refrigeration
|
No Refrigeration
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|