|
Eppendorf Thermomixer C Model 5382
Speed Range 300 to 3,000 RPM
Temperature Range 15°C to 100°C
Maximum Heating/Cooling Rate Heating: 7°C/min - Cooling: 2.5°C/min
- Provides excellent mixing performance with precise temperature control for reliable test results
- Suitable for a variety of applications including DNA/RNA purification and PCR setup mixing
- Features quick and easy-to-exchange SmartBlocks for compatibility with common lab vessels
- Offers a maximum speed of 3,000 rpm for efficient sample mixing
- Temperature control range from a minimum of 15 °C to a maximum of 100 °C
- Mixing and vortexing radius of 1.5 mm with a mixing frequency range of 300 – 3,000 rpm
- High temperature accuracy of max. ±0.5 °C between 20 – 45 °C
- Maximum heating rate of 7 °C/min (when using SmartBlock plates) and cooling rate of 2.5°C/min between 100°C and RT
The Eppendorf ThermoMixer C combines excellent mixing performance with excellent temperature control to ensure complete, dependable and reproducible test results. Improve your assay results by mixing and incubating samples at the same time. It is used for a variety of applications like plasmid/RNA/DNA purification, cDNA synthesis, transformation of bacteria strains and plasmids, mixing of PCR setups, and many more. The Eppendorf ThermoMixer C convinces users with its quick and easy-to-exchange SmartBlocks for all common lab vessels, new and improved program functions, and high level of stability for every mixing frequency.
The Eppendorf Thermomixer C Model 5382 is in excellent condition, and all of our equipment is fully tested to perform at factory specifications before being shipped.
The Eppendorf Thermomixer C Model 5382 includes:
- > 1.5 ml Block
- > ThermoTop
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Eppendorf Thermomixer C Model 5382 Specifications
Max. Speed
|
3,000 rpm
|
Max. Temperature Control
Range
|
100 °C
|
Min. Temperature Control
Range
|
15 °C
|
Mixing and Vortexing
Radius
|
1.5 mm
|
Mixing Frequency
|
300 – 3,000 rpm
|
Mixing Orbit
|
3 mm
|
Temperature Accuracy
|
Max. ±0.5 °C at
20 – 45 °C
|
Temperature Settings
|
1 °C / 100 °C
|
Temperature Range
|
Min: 15 °C below
RT, max: 100 °C
|
Dimensions (cm)
|
20.6 x 30.4 x 13.6
|
Heating Rate
Cooling Rate
|
Max. 7 °C/min (when
using SmartBlock plates)
Max. 2.5 °C/min
between 100 °C and RT
|
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stability and accuracy of power supply and PCB voltage outputs
|
Completed
|
|
Inspect display integrity, ensuring pixel clarity and replacing any degraded elements
|
Completed
|
|
Verify functionality and responsiveness of all control buttons, switches, and knobs
|
Completed
|
|
Inspect internal wiring and connections for continuity, replacing or securing degraded components
|
Completed
|
|
Examine block holder platform for alignment, obstructions, or irregularities
|
Completed
|
|
Inspect motor bearings for wear or misalignment, replacing if necessary
|
Completed
|
|
Calibrate rotational speed using an NIST-traceable tachometer
|
Completed
|
|
Calibrate temperature with an NIST-traceable probe, validating accuracy
|
Completed
|
|
Ensure timer precision by validating against set parameters
|
Completed
|
|
Conduct cosmetic inspection for quality standard compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|