Eppendorf BioPhotometer D30 Model #6133
The Eppendorf BioPhotometer D30 Model #6133 is a UV-Vis photometer for measuring liquids in cuvettes. As the measuring data are collected at fixed wavelengths, the device especially is suited for routine applications in the biomolecular, biotechnological, biochemical, and cytological areas in research and development. The single-beam photometer with reference beam and several fixed wavelengths make this photometer extremely efficient within the lab with exceptional reliability and repeatability. The high quality components guarantee high operating reliability while the metal housing ensures operating stability.
The Eppendorf BioPhotometer D30 Model #6133 includes:
- Photometric Calibration Report
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Eppendorf BioPhotometer D30 Model #6133 Specifications
|
Optical system |
Absorption single-beam photometer with reference beam and several fixed wavelengths |
|
Irradiation source |
Xenon flash lamp |
|
Spectral dispersion |
Holographic aberration-corrected concave grating |
|
Measuring wavelengths |
230, 260, 280, 320, 340, 405, 490, 562, 595, 600 nm |
|
Wavelength selection |
Method-dependent, program controlled |
|
Spectral bandwidth |
≤ 4 nm |
|
Wavelength systematic error |
± 1 nm |
|
Photometric random error |
≤ 0.002 A at 0 A ≤ 0.005 A at 1 A |
|
Photometric systematic error |
± 1% at 1 A |
|
Accuracy of reading |
0.001 A |
|
Stray-light component |
< 0.05% |
|
Memory |
> 100 method programs
|
|
Supply voltage |
100 to 240 V ± 10%: 50 to 60 Hz |
|
Dimensions (W x D x H) |
29.5 x 40 x 15 cm |
|
Weight |
11.9 lbs (5.4 kg) |
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm stable, correct voltage outputs from power supply and PCB | Completed | |
| Inspect display for dim or dead pixels; replace as needed | Completed | |
| Verify functionality of all buttons for precise, responsive control | Completed | |
| Check XY assembly for proper movement; clean railings to remove debris | Completed | |
| Confirm stable software communication between instrument and PC | Completed | |
| Inspect and clean or replace filters to maintain optimal performance | Completed | |
| Examine internal wiring for damage; replace or secure compromised connections | Completed | |
| Test lamp functionality; replace if dimming or faults are detected | Completed | |
| Execute validation protocol and document on calibration/verification certificate | Completed | |
| Update PC operating system for compatibility and security | Completed | |
| Conduct cosmetic inspection to ensure quality of external components | Completed |
Downloads