Absorbance Specifications
|
Light Source
|
Xenon flash lamp
|
Detector
|
Photodiode
|
Wavelength Selection
|
Monochromator
|
Wavelength Range
|
200 to 999 nm, in 1 nm increments
|
Wavelength Bandwidth
|
°¤ 5 nm
|
Wavelength Accuracy
|
± 2 nm
|
Wavelength Precision
|
± 0.2 nm (standard deviation)
|
Dynamic Range
|
0 to 4.0 OD
|
Photometric Resolution
|
0.0001 OD (under Gen5 control)
|
Pathlength Correction
|
Yes (under Gen5 control)
|
Photometric Accuracy
|
<1% at 2.0 OD
<3% at 2.5 OD
|
Photometric Linearity
|
< 1% from 0 to 2.5 OD
|
Photometric Repeatability
|
< 0.5% at 2.0 OD
|
Stray Light
|
0.03% at 230 nm
|
Reading Speed (Kinetic)
|
96 wells: 12 seconds
384 wells: 23 seconds
|
Fluorescence Specifications
|
Light Source
|
Halogen
|
Detector
|
PMT
|
Wavelength Selection
|
Bandpass filters
|
Wavelength Range
|
320 to 700 nm (low noise PMT)
320 to 850 nm (red-shifted PMT
|
Dynamic Range
|
> 6 decades
|
Sensitivity
|
Fluorescein 2 pM
|
Reading Speed (Kinetic)
|
96 wells: 24 seconds
384 wells: 76 seconds
|
Luminescence Specifications
|
Dynamic Range
|
< 6 decades
|
Sensitivity
|
10 amol ATP
|
General Specifications
|
Microplate Formats
|
6, 12, 24, 48, 96, 384 wells
|
Read Methods
|
End point, kinetic, area scanning, absorbance spectral scanning (under Gen5 control)
|
Plate Shaking
|
Linear, orbital, double-orbital
|
Read Time
|
96 wells: Abs 18 sec., Fl 17 sec., Lumi 120 sec.
384 wells: Abs 49 sec., Fl 48 sec., Lumi 420 sec.
|
Dimensions (H x W x D)
|
12 x 15 x 15 in. (30.5 x 38.1 x 38.1 cm)
|
Weight
|
27 lb. (12.3 kg.)
|
Electrical Requirements
|
100-240V, 50/60Hz
|