Block Configuration
|
96-Well (1 to 50 uL reactions, recommended for 10 to 25 uL reactions)
|
|
Temperature Range
|
0°C to 100°C
|
|
Temperature Uniformity
|
± 0.4°C well-to-well within 10 seconds of arrival at 90°C
|
|
Temperature Accuracy
|
± 0.2°C of programmed target at 90°C
|
|
Maximum Ramp Rate
|
5°C/second
|
|
Average Ramp Rate
|
3.3°C/second
|
|
Gradient Operational Range
|
30°C to 100°C
|
|
Gradient Programmable Span
|
1°C to 24°C
|
|
Scan Time
|
12 seconds (all channels), 3 seconds (FAM only)
|
|
Excitation Source
|
3 filtered LEDs
|
|
Detection Channels
|
3 filtered photodiodes
|
|
Range of Excitation/Emission Wavelengths
|
450 to 580 nm
|
|
Dynamic Range
|
10 orders of magnitude
|
|
Sensitivity
|
Detects 1 copy of target sequence in human genomic DNA
|
|
Multiplexing
|
Up to 2 targets per well
|
|
Dimensions (W x D x H)
|
13 x 18 x 14 in. (33 x 46 x 36 cm)
|
|
Weight
|
47 lb. (21 kg)
|
|
Electrical Requirements
|
100-240V, 50/60Hz
|