 |
 |
 |
 |
Waters ACQUITY UPLC I-Class System w/ PDA & FLR Detectors
Wavelength Range (PDA) 190 to 500 nm
Wavelength Range (FLR) EX: 200 to 890 nm / EM: 210 to 900 nm
Pressure Range Up to 18,000 psi
- UPLC platform designed for sub-2 um particle columns and ultra-high-pressure operation up to 18,000 psi
- Significantly reduced system dispersion for sharper peaks and higher chromatographic resolution
- Binary solvent manager with ballistic gradient capability for faster cycle times and high method reproducibility
- Flow-through-needle sample manager delivering high injection precision, low carryover, and efficient sample recovery
- Active column heating with mobile phase preheating for consistent thermal control and reliable method transfer
- Photodiode array detector providing full UV/Vis spectral data with high wavelength accuracy and fast acquisition rates
- Fluorescence detector optimized for trace-level analysis with high sensitivity and excellent signal-to-noise performance
- Flexible configuration supporting complex separations, dilute samples, and demanding analytical workflows
- Optimized low-volume fluidics to support high sensitivity and minimal band broadening
- Wide injection volume flexibility with optional extension loops and auto-dilution capability
- Temperature-controlled sample compartment to maintain sample integrity during analysis
- Robust system design supporting consistent, reproducible results across extended analytical sequences
The Waters ACQUITY UPLC I-Class System w/ PDA & FLR Detectors is a high-performance ultra-high-pressure liquid chromatography platform engineered for laboratories that require exceptional resolution, sensitivity, and reproducibility when analyzing complex or low-concentration samples. Built on Waters’ Ultra Performance Liquid Chromatography (UPLC) technology, the I-Class system leverages sub-2 um column particle sizes and operating pressures up to 18,000 psi to dramatically improve chromatographic efficiency compared to conventional HPLC systems. By reducing system dispersion and optimizing fluidic pathways, the platform enables sharper peak shapes, higher peak capacity, and faster separations, making it particularly well-suited for challenging analytical methods and demanding throughput requirements.
At the core of the system is the ACQUITY UPLC I-Class Binary Solvent Manager, which is specifically designed for high-pressure, low-volume UPLC operation. The binary pumping system supports precise isocratic and gradient methods across a broad flow range, with advanced pressure management that enables ballistic gradients for faster cycle times without compromising separation quality. Integrated vacuum degassing, intelligent intake valves, automated priming, and active plunger seal washing contribute to stable solvent delivery, reduced maintenance, and consistent long-term performance. These capabilities allow the system to operate within the optimal linear velocity range for sub-2 um particle columns, supporting reliable and reproducible method execution.
Sample handling is performed by the ACQUITY UPLC I-Class Sample Manager-FTN, which uses a flow-through-needle injection mechanism to deliver high precision and low carryover across a wide range of injection volumes. This design allows gradients to pass directly through the needle during injection, ensuring complete sample transfer and efficient recovery. Optional extension loops and auto-dilution functionality provide additional flexibility for varying sample concentrations and volumes, while temperature-controlled sample storage helps maintain sample integrity. The injector’s low carryover characteristics are particularly beneficial for trace-level analysis and workflows that demand high sensitivity and reproducibility.
Thermal control within the system is managed by the ACQUITY UPLC I-Class Column Heater (CH-A), which plays a critical role in maintaining consistent chromatographic conditions. By actively preheating the mobile phase and injected sample to match the column temperature, the system minimizes thermal gradients that can affect retention time and peak shape. Precise temperature control improves method robustness and supports reliable method transfer between systems, especially for applications where thermal stability is essential to achieving reproducible results.
Detection is provided by the integrated photodiode array (PDA) and fluorescence (FLR) detectors, offering complementary analytical capabilities within a single platform. The PDA detector delivers full UV/Vis spectral acquisition across a wide wavelength range with high optical resolution and wavelength accuracy, supporting peak purity assessment, compound identification, and quantitative analysis. The FLR detector adds exceptional sensitivity for compounds that exhibit native fluorescence or are derivatized for fluorescence detection. Using a high-intensity mercury-xenon arc lamp, monochromator-based wavelength selection, and a low-volume quartz flow cell, the FLR detector is optimized to maximize signal-to-noise performance while minimizing background interference.
Together, the system’s advanced pumping, injection, thermal control, and detection technologies make the Waters ACQUITY UPLC I-Class System with PDA and FLR Detectors a comprehensive solution for high-performance liquid chromatography. It is particularly well-suited for applications that demand fast analysis, high sensitivity, and consistent, reproducible data, including pharmaceutical research, bioanalysis, environmental testing, and complex chemical separations where analytical confidence and method reliability are critical.
The Waters ACQUITY UPLC I-Class System w/ PDA & FLR Detectors can be fully controlled from the Waters Empower Software, which allows for user-friendly operation in data acquisition, analysis, instrument control, automation, and more.
This Waters system is in excellent condition and will be fully tested to perform at factory
specifications before being shipped.
The Waters ACQUITY UPLC I-Class System w/ PDA & FLR Detectors includes:
- > 186015032 UV/Vis PDA Detector
- > 186015029 Fluorescence FLR Detector
- > 186015042 CH-A Column Heater
- > 186015046 SM-FTN Sample Manager
- > 186015000 BSM Binary Solvent Manager
- > Windows 11 Professional 64-Bit Data Station with Flatscreen Monitor
- > Empower Software Installed
- > Communication Cables
- > Solvent Tubing
- > 4 x Bottles, Caps and Solvent Filters
- > 1/4" Wrench & Mobile Phase Priming Syringe
- > Waste Tubing
- > All Power Cords
- > Waters Diagnostics Report
- > Full 180 Day Parts and Labor Warranty
Waters ACQUITY UPLC I-Class System w/ PDA & FLR Detectors Specifications
|
PDA Detector
|
Specification
|
|
Wavelength Range
|
190 to 500 nm
|
|
Optical Resolution
|
1.2 nm
|
|
Digital Resolution
|
1.2, 2.4, 3.6, 4.8, 6.0, 7.2, 8.4, 10.8, 12.0
|
|
Wavelength Accuracy
|
± 1.0 nm
|
|
Wavelength Repeatability
|
± 0.1 nm
|
|
Digital Filter
|
Variable with data rate
|
|
Order Filter
|
Fixed, 340 to 500 nm
|
|
Noise (Shunt in Place of Flow Cell)
|
1 - 400 nm
|
|
Noise (10 mm Analytical Flow Cell)
|
4 nm
|
|
Drift (Medium Shunt Cell and 10 mm Analytical Flow Cell)
|
~ 0.5 nm
|
|
Linearity
|
< 5% at 2.0 AU, propylparaben series at 257 nm
|
|
Data Rate
|
1, 2, 5, 10, 20, 40, and 80
|
|
FLR Detector
|
Specification
|
|
Wavelength Range
|
200 to 890 nm (Excitation)
210 to 900 nm (Emission)
|
|
Bandwidth
|
20 nm
|
|
Wavelength Accuracy
|
± 3 nm
|
|
Wavelength Repeatability
|
± 0.25 nm
|
|
Sensitivity
|
Signal-to-noise > 1000 (Raman spectrum of water)
|
|
Measurement Range
|
0.001 to 10,000 EU
|
|
Data Acquisition Rate
|
Up to 80Hz
|
|
Unattended Operation
|
Leak sensor, full 96-hour diagnostic data displayed through ACQUITY UPLC console software
|
|
Light Source
|
Hg/Xe arc lamp
|
|
Flow Cell Design
|
Axially illuminated
|
|
Cell Volume
|
< 2 uL
|
|
Pressure Limit
|
3447 kPa (34 bar, 500 psi) for standard flow cell
|
|
Materials
|
Stainless steel, fused silica, FEP, PEEK
|
Click Here to View More Specifications for the Waters ACQUITY UPLC I-Class System w/ PDA & FLR Detectors
|
BSM Binary Solvent Manager
|
Specification
|
|
Number of Solvents
|
Up to four, in combination of two, A1 or A2 and B1 or B2
|
|
Solvent Conditioning
|
Integrated vacuum degassing, six lines with two allocated for the injector needlewash/purge solvents
|
|
Gradient Formation
|
High pressure mixing, binary gradient
|
|
Gradient Profiles
|
11 gradient curves, including linear, step (2), concave (4), and convex (4)
|
|
Primary Check Values
|
Intelligent intake valves
|
|
Flow Accuracy
|
± 1.0% of set flow at 0.500 mL/min as per SystemsQT
|
|
Flow Precision
|
0.075% RSD or 0.01 min SD, (0.2 to 2.0 mL/min), whichever is greater using premixed solvent
|
|
Composition Ripple (Baseline Noise)
|
< 1.0 mAu
|
|
Compositional Precision
|
< 0.2 RSD, or 0.02 min SD, whichever is greater (from 0.2 to 2.0 mL/min)
|
|
Composition Accuracy
|
± 0.5% absolute from 5% to 95%, 0.2 to 2.0 mL/min (referenced to 100% solvent B)
|
|
Compressibility Compression
|
Automatic, no user intervention required
|
|
Priming
|
Wet priming runs at a flow rate of 4 mL/min per pump
|
|
Plunger Seal Wash
|
Equipped with a programmable active wash system, to flush the rear of the high pressure seals and plungers
|
|
Flow Ramping
|
Automatic
|
|
Mixing Options
|
Standard: 50 uL
Optional: 100 uL and 380 uL
|
|
Composition Range
|
0.0 to 100.0% settable in 0.1% increments
|
|
SM-FTN Sample Manager
|
Specification
|
|
Injection Volume Range
|
0.1 to 10.0 uL as standard configuration
Up to 1000.0 uL with optional extension loop
|
|
Injection Accuracy
|
± 0.2 uL, measured by fluid weight removed from vial with 10.0 uL injections averaged over 20 injections using standard 100 uL syringe
|
|
Injection Linearity
|
R
2 > 0.999
|
|
Injection Precision
|
< 1% area RSD 0.2 to 1.9 uL injection
< 0.5% area RSD 2.0 to 10.0 uL injection
|
|
Maximum Sample Capacity
|
Any two of the following:
96 and 384 microtiter plates 48-position 2.0 mL vial plates 48-position 0.65 mL microcentrifuge tube plates 24-position 1.5 mL microcentrifuge tube plates
|
|
Sample Compartment Temperature Range
|
4 to 40°C, settable in 0.1°C increments with a tolerance range between -2 and +4°C
|
|
Temperature Accuracy
|
No more than a ± 0.5 °C in temperature between a traceable external temperature measurement device and instrument temperature measurement device
|
|
Temperature Stability
|
± 1.0°C (at the sensor with sample compartment door closed)
|
|
Injection Needle Wash
|
Integrated, active, programmable
|
|
Minimum Sample Required
|
3 uL residual, using Waters' total recovery 2 mL vials (zero offset
|
|
Sample Carryover - UV
|
< 0.001%
|
|
Sample Carryover - MS
|
< 0.001%
|
|
CH-A Column Heater
|
Specification
|
|
Column Capacity
|
Single column, up to 4.6 mm internal
diameter (ID), up to 150 mm in length
with filter or guard column. Mounting
extends out for use with MS-based
detector
|
|
Fittings
|
124,106 kPa (1241 bar, 18,000 psi), low
dispersion, with reusable column inlet
fittings
|
|
Temperature Range
|
20 to 90°C, adjustable in increments of 0.1°C (control requires a setpoint of greater than ambient +5°C
|
|
Temperature Accuracy
|
Tested to ± 0.5°C
|
|
Temperature Stability
|
Tested to ± 0.3°C
|
|
Solvent Conditioning
|
Active pre-heating as standard
Passive pre-heating (also recommended in CH-A only for legacy method support
|
|
Column Tracking
|
eCord technology column information management tracks and archives column usage history
|
|
Computer Specifications
|
|
Processor
|
Intel Core I5-8500 CPU
|
|
RAM
|
32.0 GB
|
|
Hard Drive
|
500GB Internal SSD
|
|
Operating System
|
Windows 11 Professional, 64-bit
|
- Verify uniformity of main and residential firmware across all components, updating as necessary to ensure compatibility
- Confirm successful communication between software and the HPLC system, validating data transfer integrity
- Check the degasser’s internal vacuum to confirm it reaches the specified vacuum level for optimal performance
- Replace inlet, outlet, purge valve, and pump seals in the system pump to maintain reliable operation
- Ensure system pump passes pressure and leak tests, confirming secure and efficient performance
- Replace all capillary tubing with new OEM stainless steel tubing for durability and precision
- Remove autosampler transport assembly, replace X and Y bearings, and thoroughly clean carrier rails to remove debris and contaminants
- Replace metering valve seal, autosampler needle, and needle seat for consistent sampling accuracy
- Test column compartment switching valve functionality and replace its seal if necessary
- Confirm column compartment passes thermostat test to maintain precise temperature control
- Remove and clean internal detector optics for clear and accurate signal detection
- Replace UV and Vis detector lamps to ensure optimal light source stability and performance
- Verify flow cell absorbance meets manufacturer’s specifications for accurate measurement
- Confirm system response to isocratic test sample, ensuring stable and reproducible results
Technician Inspection and Recertification Checklist: HPLC Systems
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Verify firmware uniformity and update as necessary
|
Completed
|
|
|
Confirm software communication with HPLC system
|
Completed
|
|
|
Check degasser vacuum level for optimal performance
|
Completed
|
|
|
Replace inlet, outlet, purge valve, and pump seals
|
Completed
|
|
|
Ensure pump passes pressure and leak tests
|
Completed
|
|
|
Replace all capillary tubing with OEM stainless steel tubing
|
Completed
|
|
|
Remove autosampler transport assembly, replace bearings, and clean rails
|
Completed
|
|
|
Replace metering valve seal, autosampler needle, and needle seat
|
Completed
|
|
|
Test column compartment switching valve and replace seal if needed
|
Completed
|
|
|
Confirm thermostat test in column compartment
|
Completed
|
|
|
Clean internal detector optics
|
Completed
|
|
|
Replace UV and Vis detector lamps
|
Completed
|
|
|
Verify flow cell absorbance meets manufacturer’s specifications
|
Completed
|
|
|
Confirm system response to isocratic test sample
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|