VWR Mega Star 1.6 Centrifuge
Maximum Speed 15,200 RPM (25,830 x g RCF)
Maximum Capacity 4 x 400mL
Timer Up to 9 hours 59 minutes, and continuous run
- Compact, space-saving design with ergonomic height and curved front for easy loading and cleaning
- Intuitive keypad interface for simple control and real-time status monitoring
- Auto-Lock™ III Rotor System enables quick, push-button rotor exchange and easy chamber access
- ClickSeal™ biocontainment lids offer glove-friendly, one-handed operation with certified sealing
- Compatible with a variety of tubes and microplates for versatile lab applications
- Supports up to 40% energy savings during standard separation protocols
- Robust construction designed to meet modern safety and certification standards
- Available in individual or package formats, including a tissue culture configuration
The VWR Mega Star 1.6 Centrifuge is a high-performance general-purpose benchtop centrifuge designed to deliver consistent, secure, and reproducible results in a compact, easy-to-use platform. With a maximum capacity of 4 × 400 mL and support for a wide variety of tube formats and microplates, it’s ideal for routine applications across clinical, research, and industrial settings. Featuring an intuitive keypad interface, users can quickly set and monitor parameters such as run time, acceleration/deceleration profiles, and processing status. The unit supports up to 15,200 RPM and 25,830 × g, offering strong separation performance for a range of protocols.
Engineered with the Auto-Lock™ III Rotor System for fast rotor exchange and the ClickSeal™ Biocontainment System for simple, secure sealing, the Mega Star 1.6 enhances both user safety and lab efficiency. Its compact footprint and ergonomic design optimize bench space while enabling easy sample access and maintenance. Additionally, it offers up to 40% energy savings during standard protocols, making it an environmentally conscious choice. With up to 9 hours, 59 minutes of programmable run time and noise levels under 73 dB(A), this centrifuge ensures both flexibility and comfort during extended use. All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The VWR Mega Star 1.6 Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Guide
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
VWR Mega Star 1.6 Centrifuge Specifications
|
Maximuim Speed
|
15,200 RPM (fixed angle rotor)
|
|
Minimum Speed
|
300 RPM
|
|
Maximum RCF
|
25,830 x g
|
|
Maximum Capacity
|
4 x 400ml
|
|
Temperature Range
|
N/A
|
|
Timer
|
Up to 9 hours 59 minutes, and continuous run
|
|
Accleration Rates
|
9
|
|
Deceleration Rates
|
10
|
|
Maximum Kinetic Energy
|
41 kJ
|
|
Noise Level at Max Speed
|
< 73 dB (A)
|
|
Dimensions (H x W x D)
|
14.3 x 17.5 x 26 in. (36.2 x 44.5 x 66 cm)
|
|
Weight
|
134.5 lb. (61 kg)
|
|
Electrical Requirements
|
100-240V, 50/60Hz, 850W, 8.5A
|
|
Refrigeration
|
No Refrigeration
|
Click Here to Expand Rotor Specifications
|
TX-200 Rotor Specs
|
|
Capacity
|
4 x 180 mL bottles
|
|
Max RPM
|
5,500
|
|
Max RCF
|
5,580 x g
|
|
Fiberlite F-15-6x100Y Rotor Specs
|
|
Capacity
|
6 x 100 mL Tubes
|
|
Max RPM
|
15,000
|
|
Max RCF
|
24,652 x g
|
|
M-20 Microplate Rotor Specs
|
|
Capacity
|
2 x 3 Standard or 2 x 1 Deepwell Plates
|
|
Max RPM
|
4,000
|
|
Max RCF
|
2,272 x g
|
|
TX-400 Rotor Specs
|
|
Capacity
|
4 x 400 mL Bottles
|
|
Max RPM
|
5,000 RPM
|
|
Max RCF
|
4,696 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
| Task |
Completed? |
Remarks? |
| Confirm power supply output voltages match specified requirements |
Completed |
|
| Measure and verify voltages at PCB test points for accuracy |
Completed |
|
| Confirm rotor compatibility and detection for proper identification |
Completed |
|
| Inspect exterior for cosmetic or structural damage |
Completed |
|
| Check door mechanism alignment for secure opening and closing |
Completed |
|
| Inspect and replace gas strut with OEM part as needed |
Completed |
|
| Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
| Replace and confirm functionality of all lid lock switches |
Completed |
|
| Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
| Test compressor (if applicable) for proper operation and cooling |
Completed |
|
| Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
| Check rotor motor for stability, noise, and integrity |
Completed |
|
| Inspect and clear condensation drain to prevent buildup |
Completed |
|
| Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
| Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|