VWR Gravity Convection Incubator, Cat # 414004-620
The VWR Gravity Convection Incubator is for basic incubation with reduced air changes. With a temperature range of an ambient +8C to 65C it is accurate within 0.3C at 37C and a heating time of 30 min. to 37C and 45 min. to 50C. It provides a safe environment while reducing drying samples out. The temperature sensor allows for repeatability and stability over an extended period of time. It has a digital backlit LCD display for easy viewing. All of our equipment is fully tested to perform at factory specifications before being shipped.
Included with the VWR Gravity Convection Incubator, Cat # 414004-620:
Includes: > Shelf > Power Cord > Full 180 Day Parts and Labor Warranty
VWR Gravity Convection Incubator, Cat # 414004-620 Specifications
|
Load per Rack
|
15 kg (33 lbs.)
|
|
Permitted Total Load
|
25 kg (55 lbs.)
|
|
Wall Clearance
|
Rear: 15 cm (5.9)
Side: 15 cm (5.9")
|
|
Exhaust Duct (Outer)
|
4 cm (1.6")
|
|
Temperature Range
|
Ambient Temperature +8C to 65C
|
|
Temperature Accuracy
|
within 0.3C at 37C
|
|
Temperature setting sensitivity
|
0.1C
|
|
Temperature variation
|
0.7C at 37C, 1.2C at 50C
|
|
Heating Time
|
30 min. to 37C, 45 min. to 50C
|
|
Temperature Recovery (Door Open 30 Sec.)
|
11 min. to 37C
16 min. to 50C
|
|
Volume
|
50 L (1.75 cu. Ft)
|
|
Interior Dimensions (L x W x H)
|
54.5 54.1 70.1 cm
(21 7/16 21 5/16 27 5/8 in)
|
|
Electrical
|
120V, 60Hz, 200W
|
|
Shipping weight
|
53 kg (118 lbs.)
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|