VWR Cooling Thermal Shake Touch Cat. # 89232-904
Maximum Speed 300 to 3,000 RPM (300 to 2,200 RPM with Included Block)
Temperature Range
Ambient -17°C to 100°C
Temperature Accuracy
± 0.5°C between 20°C and 45°C
± 2°C below 20°C and above 45°C
- Intuitive 4.3" color LCD touchscreen for simple operation
- Rapid heating and cooling for precise temperature control
- Programmable memory for automated multi-step processes
- PID controller maintains consistent temperature stability
- Adjustable temperature ramp rates for customized heating/cooling
- Pulse mode feature ideal for quick vortex mixing
- Cool-touch housing enhances safety during use
- Audible alarm indicates timer completion or set temperature reached
The VWR Cooling Thermal Shake Touch Cat. # 89232-904 delivers precise temperature control and reliable shaking performance for demanding laboratory applications. It features a speed range of 300 to 3,000 RPM (300 to 2,200 RPM with standard block) with ±2% accuracy, and temperature control from 17°C below ambient to 100°C. The intuitive 4.3" color LCD touchscreen enables rapid, simultaneous adjustment of speed, temperature, and time, while clearly displaying real-time status. A PID temperature controller maintains accuracy within ± 0.5°C between 20°C and 45°C, and ± 2°C beyond this range. Heating rates up to 5°C per minute and cooling to 3°C ensure efficient operation, complemented by user-programmable memory for automated, multi-step processes. Compact footprint (10.25 x 9.75 x 5.2 in.) optimizes bench space, and advanced safety features - including cool-touch housing, limit controls, and audible alarms - ensure secure, consistent operation.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR Cooling Thermal Shake Touch Cat. # 89232-904 includes:
- > 1.5ml Block
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
VWR Cooling Thermal Shake Touch Cat. # 89232-904 Specifications
|
Speed Range
|
300 to 3,000 RPM (300 to 2,200 RPM with Included Block)
|
|
Speed Accuracy
|
± 2%
|
|
Temperature Range
|
Ambient -17°C to 100°C
|
|
Temperature Accuracy
|
± 0.5°C between 20°C and 45°C
± 2°C below 20°C and above 45°C
|
|
Heating Rate
|
5°C/minute
|
|
Cooling Rate
|
Above Ambient: 2 to 3°C/minute
Below Ambient: 0.5 to 1°C/minute
|
|
Timer
|
1 minute to 99 hours, 59 minutes
|
|
Orbit
|
3mm (0.12 in.)
|
|
Dimensions (L x W x H)
|
10.25 x 9.75 x 5.2 in. (26.04 x 24.77 x 13.2 cm)
|
|
Weight
|
17 lb. (7.7 kg)
|
|
Electrical Requirements
|
120V, 50/60Hz, 1.8A, 215W
|
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Confirm stability and accuracy of power supply and PCB voltage outputs
|
Completed
|
|
|
Inspect display integrity, ensuring pixel clarity and replacing any degraded elements
|
Completed
|
|
|
Verify functionality and responsiveness of all control buttons, switches, and knobs
|
Completed
|
|
|
Inspect internal wiring and connections for continuity, replacing or securing degraded components
|
Completed
|
|
|
Examine block holder platform for alignment, obstructions, or irregularities
|
Completed
|
|
|
Inspect motor bearings for wear or misalignment, replacing if necessary
|
Completed
|
|
|
Calibrate rotational speed using an NIST-traceable tachometer
|
Completed
|
|
|
Calibrate temperature with an NIST-traceable probe, validating accuracy
|
Completed
|
|
|
Ensure timer precision by validating against set parameters
|
Completed
|
|
|
Conduct cosmetic inspection for quality standard compliance
|
Completed
|
|