VWR Analog Standard Heat Block III, Cat. #: 13259-034
Temperature Range Ambient +5° to 150°C
Temperature Stability ± 2°C at 37°C (± 4°C at 60°C)
Temperature Uniformity ± 0.1°C at 37°C (± 0.6°C at 60°C)
- Ideal for activation of cultures, enzyme reactions, and immunoassays
- Versatile performance suitable for melting and boiling point determinations
- Heat indicator light provides clear visual confirmation of heater operation
- Compact, low-profile design conserves valuable bench space
- Reliable temperature consistency ideal for a wide variety of laboratory applications
The VWR Analog Standard Heat Block III, Cat. #: 13259-034 accommodates 3 blocks and features a powerful heater with uniform temperature control in sample processing. This model accepts several interchangeable blocks to ensure versatility across research applications. The VWR Analog Standard Heat Block III, Cat. #: 13259-034 integrates a heating light into the display to notify users when the heating function is on, providing safety within laboratory settings. Its precise and multi-purpose functionality makes this heat block ideal for a number of applications including activation of cultures, enzyme reactions, immunoassays and more.
Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire.
All of our equipment is fully tested before being shipped.
The VWR Analog Standard Heat Block III, Cat. #: 13259-034 includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
VWR Analog Standard Heat Block III, Cat. #: 13259-034 Specifications
|
Temperature Range
|
Ambient +5° to 150°C
|
|
Temperature Stability
|
± 2°C at 37°C (± 4°C at 60°C)
|
|
Temperature Uniformity
|
± 0.1°C at 37°C (± 0.6°C at 60°C)
|
|
Dimensions (L x W x H)
|
18.4 x 8 x 3.5 in. (46.7 x 20.3 x 8.9 cm)
|
|
Weight
|
8 lbs. (3.6 kg)
|
|
Electrical Requirements
|
120VAC, 50/60Hz, 200W
|
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Verify the number and sizes of heat blocks
|
Completed
|
|
|
Test functionality of displays, knobs, and switches
|
Completed
|
|
|
Inspect internal wiring for damage
|
Completed
|
|
|
Clean internal insulation and chamber
|
Completed
|
|
|
Check machine feet for proper balance
|
Completed
|
|
|
Confirm heat within range using an independent probe
|
Completed
|
|
|
Validate temperature uniformity across heat blocks
|
Completed
|
|
|
Inspect line cord for damage
|
Completed
|
|
|
Ensure cosmetic condition is acceptable
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|