VWR 1545 Digital Incubator
Temperature Range Ambient +5° to 70°C
Temperature Uniformity ± 0.25°C at 37C
Volume 164 L (5.8 cu. ft.)
- Digital control panel with easy-to-use temperature adjustment and calibration
- Overtemperature protection thermostat ensures system reliability
- Heating and overtemperature indicator lights provide real-time system status
- Durable stainless steel interior for easy cleaning and longevity
- Adjustable shelving system for flexible sample placement
- Insulated chamber with high-density urethane foam for energy efficiency
- Front-mounted power switch for convenient operation
- Ventilated design ensures uniform heat distribution
- Inner glass door for easy viewing of samples without disturbing chamber
The VWR 1545 Digital warm air incubator provides superior temperature
uniformity through heating elements in three walls, bottom, and door, creating
a warm air heat jacket with no hot spots. This jacket design also allows for
quick temperature recovery. The VWR 1545 features a full-length inner glass
door for easy viewing of specimens without disturbing the chamber. This
incubator is microprocessor-controlled to ensure easy keypad control and
calibration. The temperature display can be calibrated to match a secondary thermometer.
The 1545 includes a secondary independent over-temperature protection feature
as well as a through-wall access port for chart recorder.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The VWR 1545 Digital Incubator includes:
- > Calibration
- > 2 Shelves
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
VWR 1545 Digital Incubator Specifications
|
Temperature Range
|
Ambient +5° to 70°C
|
|
Temperature Uniformity
|
± 0.25°C at 37C
|
|
Volume
|
164 L (5.8 cu. ft.)
|
|
Exterior Dimensions (W x D x H)
|
25 x 27 x 36 in
|
|
Interior Dimensions (W x D x H)
|
20 x 20 x 26 in
|
|
Electrical
|
115V, 650W
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|