Thermo Scientific Steri Cycle 380 CO2 Incubator
Temperature Uniformity ± 0.3°C at 37°C
Temperature Range Ambient +5°C to 50°C
Interior Volume 6.5 cu. ft. (184L)
- Stable, accurate CO₂ control with IR sensor for critical culture protection
- High-temperature decontamination for convenience
- Temperature retention and quick recovery for critical cell cultures
- Maximum thermal protection and quick recovery from ambient temperature swings and power variations
- Patented in-chamber HEPA air filtration system for Class 100 cleanroom air quality
- 140°C sterilization cycle ensures high-level decontamination with minimal user effort
- Interactive LED control panel with multi-mode navigation for streamlined setup and monitoring
The Thermo Scientific™ Steri-Cycle 380 CO₂ Incubator delivers advanced contamination control and precise environmental stability for sensitive cell cultures. Designed with a reliable infrared (IR) CO₂ sensor and a direct-heat chamber, this 6.5 cu. ft. incubator ensures accurate, stable gas control and rapid recovery from temperature fluctuations or door openings. Its integrated 140°C sterilization cycle simplifies routine decontamination with a 12-hour, user-initiated process that reduces contamination risk between studies.
An interactive LED control panel provides multi-mode navigation for easy setup and monitoring, while audio/visual alarms and access code protection support secure operation. Built for durability and cleanliness, the incubator features a polished stainless-steel interior, removable shelves, and accessible microbiological filters on all gas and sample ports. With consistent performance and user-focused controls, the Steri-Cycle 380 is an ideal solution for maintaining optimal conditions in critical lab workflows.
All of our equipment is fully tested to perform
at factory specifications before being shipped.
The Thermo Scientific Steri Cycle 380 CO2 Incubator includes:
- > 3 Shelves
- > Humidity Pan
- > New Hepa Filter
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Steri Cycle 380 CO2 Incubator Specifications
|
Temperature Range
|
Ambient +5°C to 50°C
|
|
Temperature Uniformity
|
± 0.3°C at 37°C
|
|
CO2 Range
|
1 to 20%
|
|
CO2 Sensor
|
IR
|
|
Inlet Pressure
|
15psig (1 bar), ± 5 psig (0.3 bar)
|
|
Relative Humidity
|
95% RH at 37°C
|
|
Volume
|
6.5 cu. ft. (184L)
|
|
Sterilization Cycle Temperature
|
140°C
|
|
Sterilization Cycle Duration
|
12 Hours Approx.
|
|
Shelf Dimensions (W x D)
|
18.5 x 18.5 in. (47 x 47 cm)
|
|
Internal Dimensions (W x H x D)
|
21.3 x 26.8 x 20.0 in. (54.1 x 68.1 x 50.8 cm)
|
|
External Dimensions (W x H x D)
|
26.1 x 39.5 x 25 in. (54 x 68.1 x 50.8 cm)
|
|
Weight
|
260 lb. (118 kg.)
|
|
Electrical Requirements
|
115V, 50/60Hz, 1PH, 9.6 FLA
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|