Thermo Scientific Sorvall RC 12BP Plus Refrigerated Centrifuge
Temperature Range -10°C to 40°C
Maximum Capacity 6 x 2000ml
- Microprocessor-controlled system with 15 program storage capacity
- High-performance induction drive ensures reliable operation
- Eleven acceleration and deceleration profiles for optimized runs
- Noise level below 65 dB(A) for quiet laboratory operation
- Advanced safety features including automatic lid lock and imbalance detection
- Secure key switch prevents unauthorized operation
- Rugged armored guard ring enhances durability and protection
- Compact design with low heat output for in-lab placement
The Thermo Scientific Sorvall RC 12BP Plus Refrigerated Centrifuge is designed to deliver reliable and versatile performance for high-throughput, large-volume applications in blood banking and bioprocessing. This low-speed, 12-liter centrifuge operates at a maximum speed of 4,700 RPM ensuring efficient processing of substantial sample volumes at constant temperatures ranging from -10°C to 40°C. Its maintenance-free induction motor provides quiet, low-vibration operation and exceptional longevity. The user-friendly control panel of the Sorvall RC 12BP Plus allows easy presetting of speed, RCF, temperature, and running profiles, with the flexibility to adjust settings even during operation. It features a robust safety system including a high-quality steel construction, electronic imbalance detection, and an automated lid-lock that ensures the lid can only be opened when the rotor has completely stopped. The lid also includes a view port and emergency release for added safety.
Additionally, the centrifuge supports versatile bioprocessing applications with its array of compatible blood bags, tubes, and bottles. The intuitive interface, coupled with a 15-program memory, simplifies set-up and operation, while the enhanced display allows for straightforward monitoring of conditions. The Accumulated Centrifugal Effect (ACE) function ensures reproducible yields by adjusting run times based on rotor loading. This feature, along with customizable acceleration and deceleration profiles, maximizes blood product yields and minimizes re-suspension. Centri-Log data collection software enhances this centrifuge's appeal by providing comprehensive run parameter traceability essential for maintaining cGMP, GMP, and GLP compliance. The Sorvall RC 12BP Plus also boasts a compact design with minimal heat output, making it an ideal, comfortable choice for in-lab placement. Its versatility extends to processing bacterial and yeast cultures, clarifying broths, and harvesting mammalian cell and tissue cultures, facilitated by optional adapters for diverse sample vessels All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Sorvall RC 12BP Plus Refrigerated Centrifuge includes:
- > H-12000 Swing Bucket Rotor & Buckets
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Sorvall RC 12BP Plus Refrigerated Centrifuge Specifications
Maximum Speed
|
4,700 RPM
|
Minimum Speed
|
300 RPM
|
Maximum RCF
|
7,340 x g
|
Maximum Capacity
|
6 x 2000ml
|
Temperature Range
|
-10°C to 40°C
|
Timer
|
1 minute to 99 minutes 59 seconds, or continuous operation
|
Acceleration Rates
|
11
|
Deceleration Rates
|
11
|
Control
|
Microprocessor
|
Program Storage
|
15 programs
|
Noise Level
|
< 65 dB(A)
|
Drive
|
High-performance induction
|
Safety Features
|
Automatic lid lock, imbalance detection, armored guard ring, security key switch
|
Dimensions (H x W x D)
|
46.3 x 31.5 x 35.6 in. (117.8 x 80 x 90.5 cm)
|
Weight
|
1047 lb. (475 kg)
|
Electrical Requirements
|
208-220V, 60Hz, 24A, 4400W
|
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|