Thermo Scientific NanoDrop One Microvolume Spectrophotometer
Wavelength Range 190 to 850 nm
Wavelength Accuracy ± 1 nm
- Acclaro Sample Intelligence Technology identifies contaminants and ensures data integrity
- Extended auto-range pathlength enables accurate readings up to 27,500 ng/µL (dsDNA)
- High-resolution 7-inch touchscreen interface with intuitive standalone operation
- No consumables required thanks to the direct pipetting pedestal design
- Fast 8-second measurement time with auto-measure and preloaded workflows
- Xenon flash lamp and 2048-element CMOS sensor for precise UV-Vis analysis
- Multiple connectivity options including USB, Ethernet, Wi-Fi, and Bluetooth with 32 GB internal memory
The Thermo Scientific NanoDrop One Microvolume Spectrophotometer is an advanced, compact UV-Vis instrument designed to deliver fast, reliable quantification of nucleic acids, proteins, and other biomolecules using as little as 1-2 µL of sample. With its integrated Thermo Scientific Acclaro Sample Intelligence Technology, the NanoDrop One provides enhanced sample analysis through real-time contaminant identification, corrected concentration values, and immediate feedback on sample quality. These capabilities help users make confident decisions before moving to downstream applications. The instrument features a sleek, standalone design with a 7-inch high-definition touchscreen interface, eliminating the need for a connected PC and maximizing bench space efficiency.
The xenon flash lamp and 2048-element CMOS sensor provide fast, accurate results with a spectral resolution of ≤ 1.8 nm and a measurement time of just 8 seconds. Built-in Auto-Measure functionality and pre-programmed applications simplify routine workflows. For data management and transfer, the instrument includes three USB-A ports, Ethernet, Bluetooth, Wi-Fi connectivity, and 32 GB of internal flash memory. The NanoDrop One is the ideal choice for researchers requiring high-performance, intelligent microvolume analysis with the added flexibility of cuvette-based capabilities.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Scientific NanoDrop One Microvolume Spectrophotometer includes:
- > PV-1 Calibration Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific NanoDrop One Microvolume Spectrophotometer Specifications
|
Wavelength Range
|
190 to 850 nm
|
|
Wavelength Accuracy
|
± 1 nm
|
|
Photometric Range
|
0 to 550 A
|
|
Photometric Accuracy
|
3% at 0.97 A, 302 nm
|
|
Minimum Sample Size
|
1 uL
|
|
Pathlength
|
0.030 to 1.0 mm auto-ranging
|
|
Light Source
|
Xenon flash lamp
|
|
Detector
|
2048-element CMOS linear image sensor
|
|
Resolution (Spectral Bandwidth)
|
≤ 1.8 nm (FWHM at Hg 254 nm)
|
|
Lower Limit of Detection
|
2.5 ng/uL (dsDNA)
0.15 mg/mL (BSA)
|
|
Maximum Concentration
|
27,500 ng/uL (dsDNA)
820 (400) mg/mL (BSA lgG)
|
|
Measurement Time
|
8 seconds
|
|
Connectivity
|
Three USB-A ports, ethernet, Bluetooth, and Wi-Fi
|
|
Display
|
7 inch touchscreen, 1280 x 800 high-definition color display
|
|
Internal Storage
|
32 GB flash memory
|
|
Dimensions (W x D x H)
|
8 x 10 x 12.7 in. (20 x 25.4 x 32.3 cm)
|
|
Weight
|
7.9 lb. (3.6 kg.)
|
|
Electrical Requirements
|
12V DC, 12-18W
|
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|