Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer
Stirrer Speed 60 to 1,200 RPM
Platform Size 7 x 7 in (17.7 x 17.7 cm)
- Durable ceramic top plate offers excellent chemical resistance
- Maintains stable performance for heating and stirring applications
- Ideal for sample preparation, reagent heating, and solvent evaporation
- Compact, low-profile design maximizes bench space efficiency
- Accurate surface temperature control via internal sensor
- Reliable construction ensures long-term laboratory use
The Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer delivers precise, reliable performance for a wide range of laboratory heating and mixing applications. Featuring a durable 7 × 7 in. ceramic top plate, it ensures uniform heat distribution and excellent chemical resistance. Designed for accuracy and control, the Cimarec maintains consistent surface temperatures up to 540°C and provides smooth stirring from 60 to 1,200 rpm for volumes up to 4 liters.
Equipped with a bright digital display, this unit continuously monitors actual temperature for accurate results, while its simple interface allows easy adjustment of heating and stirring parameters. The Cimarec SP131325Q is ideal for general-purpose heating, sample preparation, reagent mixing, solvent evaporation, and media preparation. Its compact, low-profile design ensures stability and efficient bench space use, while the ceramic surface and robust construction make it dependable for daily laboratory operation.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer
Specifications
|
Maximum Temperature
|
540C
|
|
Stirrer Speed
|
60 to 1,200 RPM
|
|
Max. Stirring Volume
|
4L
|
|
Platform Size
|
7 x 7 in (17.7 x 17.7 cm)
|
|
Plate Material
|
Ceramic
|
|
Weight
|
11.0 lb. (5.0 kg)
|
|
Dimensions (W x D x H)
|
8.2 x 13 x 3.8 in. (20.8 x 33 x 9.7 cm)
|
|
Electrical Requirements
|
120V, 60Hz, 8.9A
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
| Task |
Completed? |
Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and CO₂ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|