Thermo Scientific Cimarec Plus SP88857100 Hot Plate Stirrer
Maximum Temperature 540°C
Stirrer Speed 50 to 1,500 RPM
Platform Size 7 x 7 in (17.7 x 17.7 cm)
- Advanced stirring system ensures consistent speed control and strong magnetic coupling
- Ceramic top plate provides excellent chemical resistance and durability
- Temperature adjustable in 1 degree increments for precise control
- Easy-to-read LED display shows heating and stirring parameters
- Raised control panel design helps protect electronics from spills
- HOT TOP warning system alerts users when surface is hot
- Compact, stable design supports reliable benchtop operation
- Suitable for a wide range of routine laboratory heating and stirring applications
The Thermo Scientific Cimarec Plus SP88857100 Hot Plate Stirrer is designed to deliver precise and reliable heating and stirring for routine laboratory protocols. Built with advanced engineering, this unit provides consistent speed control, strong magnetic coupling, and dependable performance even at low stirring speeds, making it well suited for a wide range of applications from basic sample preparation to more demanding workflows. The ceramic top plate offers excellent chemical resistance and durability, while the adjustable temperature control allows users to fine-tune settings in 1 degree increments for accurate and repeatable results.
An easy-to-read LED display clearly shows both heating and stirring parameters, while the raised control panel design helps protect internal electronics from spills. For added safety, the integrated HOT TOP warning system alerts users when the surface temperature exceeds 50C, reducing the risk of accidental contact. With a compact footprint and stable construction, the Thermo Scientific Cimarec Plus SP88857100 Hot Plate Stirrer provides dependable performance, user-friendly operation, and enhanced safety for everyday laboratory use. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Cimarec Plus SP88857100 Hot Plate Stirrer
includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Cimarec Plus SP88857100 Hot Plate Stirrer Specifications
|
Maximum Temperature
|
540°C
|
|
Stirrer Speed
|
50 to 1,500 RPM
|
|
Max. Stirring Volume
|
4L
|
|
Platform Size
|
7 x 7 in (17.7 x 17.7 cm)
|
|
Plate Material
|
Ceramic
|
|
Weight
|
9.0 lb. (4.1 kg)
|
|
Dimensions (W x D x H)
|
8.2 x 14.2 x 4.1 in. (20.8 x 36 x 10.5 cm)
|
|
Electrical Requirements
|
100-120V, 50/60Hz, 13A
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
| Task |
Completed? |
Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and CO₂ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|