Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer
Stirrer Speed 50 to 1,500 RPM
Platform Size 10.25 x 10.25 in (26.7 x 26.7 cm)
- Temperature adjustable in precise 1-degree increments for accurate control
- HOT TOP warning system alerts users when the surface exceeds 50C to prevent accidental burns
- Automatic over-temperature protection enhances safety during operation
- Durable ceramic top plate offers uniform heat distribution
- Easy-to-read LED display provides clear visibility of heating and stirring settings
- Raised display design safeguards electronics from spills and splashes
The Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer offers a wide speed range of 50 to 1,500 RPM accommodating a wide variety of stirring applications. This hotplate stirrer is additionally suited for samples up to 6L in volume and is capable of reaching a maximum temperature of 400°C. This heating functionality is paired with a HOT TOP warning system for safety, notifying users when the heating surface has exceeded 50°C. The simple display provides independent control of temperature and speed function, allowing users to increase each in 1°C and 10 RPM increments respectively. Designed with safety and simplicity in mind, the Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer provides an excellent general purpose option in sample processing, heating reagents, media preparation and more.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer
includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Cimarec Plus SP88850100 Digital Hot Plate Stirrer
Specifications
|
Maximum Temperature
|
400C
|
|
Stirrer Speed
|
50 to 1,500 RPM
|
|
Max. Stirring Volume
|
6L
|
|
Platform Size
|
10.25 x 10.25 in (26.7 x 26.7 cm)
|
|
Plate Material
|
Ceramic
|
|
Weight
|
13.0 lb. (5.9 kg)
|
|
Dimensions (W x D x H)
|
11.3 x 17.2 x 4.1 in. (28.8 x 43.8 x 10.5 cm)
|
|
Electrical Requirements
|
100-120V, 50/60Hz, 13.2A
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
| Task |
Completed? |
Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and CO₂ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|