Scientific Industries SI-1401 Incubator-Genie Shaking Incubator
The Scientific Industries SI-1401 Incubator-Genie Shaking
Incubator offers rotating and shaking capabilities with
temperatures from 28C to 75C for applications in gel and blot staining, cell or
bacterial culture, or any application that requires mixing, washing, rocking,
and temperature control of a sample. Magnetic platform and clip plate system
can accommodate almost any container at any angle. Programmable timers and
alarms for temperature and/or mixing parameters protect samples from deviation.
Chamber air is continuously gently recirculated to ensure temperature
uniformity.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Scientific Industries SI-1401 Incubator-Genie Shaking Incubator includes:
- > Rotating Platform
- > Calibration Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Scientific Industries SI-1401 Incubator-Genie Shaking Incubator Specifications
Rotating Speed
|
1 to 35 RPM
|
Rocking Speed
|
6 to 70 cycles per minute
|
Temperature Range
|
28C to 75C
|
Temperature Stability
|
0.1C
|
Temperature Uniformity
|
0.5C
|
Temperature Resolution
|
1.0C
|
Timer
|
Programmable to 99 hours
|
Chamber Volume
|
24 L (0.86 cu ft)
|
Platform Capacity
|
4.5 kg (10 lbs)
|
Interior Dimensions (D x W x H)
|
254 x 362 x 260 mm (10 x 14.25 x 10.25 in)
|
Overall Dimensions (D x W x H)
|
400 x 560 x 270 mm (15.75 x 22 x 14.5 in)
|
Electrical Requirements
|
230V, 50Hz, 2.5A
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|