Ohaus FC5718 Frontier Multi-Pro Centrifuge
Speed Range 200 to 18,000 RPM
Maximum Capacity 4 x 100ml
- Automatic rotor identification makes switching between applications extremely easy
- Simple one-hand operation enabled by the selection button/rotary knob combination—even with gloves on
- Large backlit LCD display shows centrifugation settings and real-time parameters on one interface
- Program storage up to 99 runs including parameter settings and rotor types
- Maintenance-free induction drive
- Thread-free rotor locking mechanism
- Optimized airflow system to keep samples cool (non-refrigerated models)
- Chemical and corrosion-resistant stainless-steel housing ensures durability in lab environments
- Self-diagnostic program to record usage and error history for easy maintenance and to minimize downtime
- Over-speed protection, immediately upon rotor insertion
- Active imbalance-detection with individual imbalance tolerances set for each rotor
- Motorized lid-locking system to protect operators from accidental injuries
The Ohaus FC5718 Frontier Multi-Pro Centrifuge is a versatile, high-performance benchtop centrifuge built to meet the rigorous demands of modern laboratories. Designed with both efficiency and safety in mind, it features a motorized lid-lock system, automatic rotor identification, and active imbalance detection—ensuring each run starts safely and ends reliably. Its large backlit LCD display clearly shows real-time run parameters, and intuitive controls with a rotary knob make adjustments simple, even when wearing gloves.
With space for up to 4 x 100 mL and support for a broad range of angle and swing-out rotors, the FC5718 adapts easily to a variety of protocols. Program storage for up to 99 runs, 10 acceleration/deceleration settings, and Quickspin and continuous run modes provide the flexibility required for both routine and specialized applications. Its maintenance-free induction drive and thread-free rotor locking system streamline daily use, while the chemical-resistant stainless steel housing ensures long-lasting durability in demanding lab environments. Compatible with sealed and biocontainment rotors, the FC5718 also provides protection from hazardous aerosols, making it a reliable and safe choice for high-throughput centrifugation needs. All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
Marshall Scientific is an Authorized Distributor of Ohaus Products. If the included 30314831 6 x 50ml Fixed Angle Rotor does not meet your requirements, our team can assist in finding a compatible brand new alternative.
The Ohaus FC5718 Frontier Multi-Pro Centrifuge includes:
- > 30314831 6 x 50ml Fixed Angle Rotor
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Ohaus FC5718 Frontier Multi-Pro Centrifuge Specifications
|
Speed Range
|
200 to 18,000 RPM (12,000 RPM w/ included Rotor)
|
|
Maximum RCF
|
23,542 x g (13,522 x g w/ included Rotor)
|
|
Maximum Capacity
|
4 x 100ml
|
|
Temperature Range
|
N/A
|
|
Timer
|
Continuous; Quickspin; 10 sec to 99 h 59 min 59 sec
|
|
Acceleration Rates
|
10
|
|
Deceleration Rates
|
10
|
|
Number of Compatible Rotors
|
18
|
|
Drive System
|
Maintenance Free Induction Drive
|
|
Lid Lock
|
Motor Driven
|
|
Display
|
LCD with Backlight
|
|
Program Storage
|
99 Programs
|
|
Noise Level
|
≤ 60 ± 2 dB(A)
|
|
Dimensions (W x D x H)
|
16.1 x 19.65 x 13.82 in (40.8 x 49.9 x 35.1 cm)
|
|
Weight
|
95 lb. (43 kg)
|
|
Electrical Requirements
|
120V, 50/60Hz, 4.0 A, 475W
|
|
Refrigeration
|
No Refrigeration
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly

- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
| Task |
Completed? |
Remarks? |
| Confirm power supply output voltages match specified requirements |
Completed |
|
| Measure and verify voltages at PCB test points for accuracy |
Completed |
|
| Confirm rotor compatibility and detection for proper identification |
Completed |
|
| Inspect exterior for cosmetic or structural damage |
Completed |
|
| Check door mechanism alignment for secure opening and closing |
Completed |
|
| Inspect and replace gas strut with OEM part as needed |
Completed |
|
| Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
| Replace and confirm functionality of all lid lock switches |
Completed |
|
| Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
| Test compressor (if applicable) for proper operation and cooling |
Completed |
|
| Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
| Check rotor motor for stability, noise, and integrity |
Completed |
|
| Inspect and clear condensation drain to prevent buildup |
Completed |
|
| Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
| Check centrifuge error log for zero recorded issues |
Completed |
|