New Brunswick Innova 4200 Incubator Shaker
Temp. Range Ambient +5°C to 80°C
Speed Range
25 to 400 RPM
Temperature Uniformity
0.3C
- Versatile incubator shaker designed for microorganism culture, cell culture, and hybridization
- Automatic interlock door shuts off shaker when the door is open for added safety
- Audible and visual alarms alert users when temperature or speed goes out of range
- Triple eccentric counterbalanced drive provides smooth and consistent shaking motion
- Horizontal plane rotary motion with a 0.75-inch circular orbit for uniform sample agitation
- Large 18 x 18 inch platform holds up to 18 x 250mL or 2 x 6L flasks
- Programmable timer allows operation from 0.1 to 99.9 hours with an audible and visual alarm
- Stackable design accommodates up to two units for space-saving configurations
The New Brunswick Innova 4200 Incubator Shaker is a versatile shaker designed for microorganism culture, cell culture, and hybridization. Automatic features include an interlock door that shuts off the shaker when opened, a chamber temperature shutoff, and an overload/unbalanced sensor that limits speed in unstable conditions. Audible and visual alarms alert users when temperature or speed deviates from set parameters, ensuring experiment integrity. The unit supports shaking speeds from 25 to 400 RPM with a precise 0.75-inch orbit, allowing flexibility for various culturing needs. Its precision temperature uniformity of ±0.3°C ensures consistent incubation conditions, while the large 18 x 18-inch platform accommodates up to 18 x 250mL or 2 x 6L flasks. A built-in timer enables operation for up to 99.9 hours, making it ideal for long-term studies. Compact chamber dimensions of 20.5 x 20.5 x 24.8 inches allow it to fit in various lab setups, while its durable construction supports reliable performance for demanding applications.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 4200 Incubator Shaker includes:
- > 18 x 18 Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
New Brunswick Innova 4200 Refrigerated Incubator Shaker Specifications
Temperature Range
|
Ambient +5C to 80C
|
Speed Range
|
25 to 400 RPM
|
Orbit Diameter
|
0.75 in
|
Temperature Uniformity
|
0.3C
|
Timer
|
Up to 99.9 hours
|
Maximum Capacity
|
18 x 250mL
2 x 6L
|
Platform Size
|
18 x 18 in
|
Chamber Dimensions (W x L x H)
|
20.5 x 20.5 x 24.8 in
|
Overall Dimensions (W x L x H)
|
24.4 x 28.7 x 36.4 in
|
Weight
|
270 lbs
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|