Memmert ICO50 CO2 Incubator
Temperature Range Ambient +5°C to 50°C
- Expert precision and control for optimal protection of cultures
- User-friendly, intuitive operating concept for ease of use
- High-quality, corrosion-resistant stainless steel construction for durability and cleanliness
- Easily cleanable inner chamber and housing
- Rapid heating to setpoint without overshoots for consistent results
- Rounded interior edges for simplified, thorough cleaning
- Interior sterilization possible at 180°C for 60 minutes
- Integrated data logger and adjustable digital timer from 1 minute to 99 days
- SetpointWAIT function ensures process timing starts only when set temperature is reached
- Battery-buffered ControlCOCKPIT ensures continuous CO2 control even during power outages
The Memmert ICO50 CO2 Incubator 56L, 115V offers unparalleled precision and control for demanding laboratory applications, ensuring optimal protection for cell cultures, bacteria cultures, and tissue cultures. Designed with user-friendliness in mind, it features an intuitive operating system complemented by a high-quality stainless steel inner chamber and housing that are both corrosion-resistant and easy to clean. This incubator excels in maintaining a hygienic environment with its rounded interior edges and the ability to sterilize the interior at 180 °C for 60 minutes without removing sensors, promoting effortless cleaning and maintenance.
Equipped with an advanced control system, the ICO50 reaches desired temperatures swiftly without overshoots, and it includes a battery-buffered ControlCOCKPIT for continuous CO2 control, even during power outages. Alarm notifications can be sent directly to your mobile device, ensuring the safety and integrity of the chamber's contents. The incubator comes with an integrated data logger and a digital timer that can be set from 1 minute to 99 days and 23 hours, supporting complex ramping processes and precise temperature control.
The SetpointWAIT function ensures that process times start only when the target temperature is reached at all monitoring points. With its robust construction and high-end safety features, the Memmert ICO50 CO2 Incubator is an indispensable tool for any research facility focused on cell and tissue culture.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Memmert ICO50 CO2 Incubator includes:
- > 1 x Stainless Steel Perforated Shelf
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Memmert ICO50 CO2 Incubator Specifications
|
Temperature Range
|
Ambient +5°C to 50°C
|
|
Temperature Uniformity
|
± 0.3°C at 37°C
|
|
Sterilization Program
|
60 minutes at 180°C (without removing the sensors)
|
|
Temperature Setting Resolution
|
0.1°C
|
|
Temperature Stability
|
± 0.1°C
|
|
Volume
|
56L (2 cu. ft.)
|
|
Humidity Control (Standard)
|
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
|
|
CO2 Range
|
0 to 20%
|
|
CO2 Control
|
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
|
|
CO2 Setting Accuracy
|
0.1%
|
|
CO2 Stability
|
± 0.2%
|
|
Control Technology
|
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays.
|
|
Language Setting
|
German, English, Spanish, French, Polish, Czech, Hungarian
|
|
Function SetpointWAIT
|
The process time does not start until the set temperature is reached
|
|
Adjustable Parameters
|
Temperature (Celsius or Fahrenheit), CO2, Timer
|
|
Communication Interfaces
|
Ethernet LAN, USB
|
|
Documentation Program
|
Stored in case of power failure
|
|
AutoSAFETY
|
Additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
|
|
Autodiagnostic System
|
Integral fault diagnostics for temperature and CO2
|
|
Alarm
|
Visual and acoustic
|
|
Heating Concept
|
6 sides large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
|
|
Door
|
fully insulated stainless steel door with 2-point locking (compression door lock)
|
|
Shelving
|
1 perforated stainless steel shelf
|
|
Inner Glass Door
|
with opening (8 mm Ø) to take gas sample
|
|
Interior Material
|
1.4301 (ASTM 304), corrosion resistant
|
|
Max. Number of Shelves
|
5
|
|
Max. Loading of Chamber
|
165 lb. (74.9 kg)
|
|
Max. Loading per Shelf
|
33 lb. (15 kg)
|
|
Internal Dimensions (W x H x D)
|
15.7 x 16.7 x 13 in. (39.9 x 42.4 x 33 cm)
|
|
External Dimensions (W x H x D)
|
22 x 31.3 x 20.5 in. (55.9 x 79.5 x 52.1 cm)
|
|
Weight
|
122 lb. (55.3 kg)
|
|
Electrical Requirements
|
115V, 50/60Hz approx., 9.6A, 1100W
|
- Verify power supply and PCB outputs are within specified voltage ranges
- Inspect display for any dim or dead pixels, replacing as needed to ensure visibility
- Confirm full functionality and responsiveness of all control buttons
- Examine all internal wiring for frays, cuts, or damage, ensuring secure and reliable connections
- Inspect door seals and gaskets for integrity, replacing any compromised components
- Evaluate door handle and locking mechanism for wear or issues, replacing if performance is impaired
- Inspect and clean internal fan; replace if performance does not meet requirements
- Calibrate set temperature using a NIST-traceable probe for precise accuracy
- Confirm functionality of over-temperature alarm and indicator light
- Verify internal outlet functionality, if applicable, to ensure it meets operational standards
- Inspect internal lighting for full functionality and brightness, if applicable
- Perform a cosmetic inspection to confirm the oven’s exterior and interior meet quality standards
Technician Inspection and Recertification Checklist: Ovens
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Verify power supply and PCB outputs are within specified voltage ranges
|
Completed
|
|
|
Inspect display for dim or dead pixels; replace if needed
|
Completed
|
|
|
Confirm full functionality of all control buttons
|
Completed
|
|
|
Examine internal wiring for damage; ensure secure connections
|
Completed
|
|
|
Inspect door seals and gaskets, replacing any compromised components
|
Completed
|
|
|
Evaluate door handle and locking mechanism, replacing if worn
|
Completed
|
|
|
Inspect and clean internal fan; replace if performance is insufficient
|
Completed
|
|
|
Calibrate set temperature using a NIST-traceable probe
|
Completed
|
|
|
Confirm over-temperature alarm and indicator light functionality
|
Completed
|
|
|
Verify internal outlet functionality, if applicable
|
Completed
|
|
|
Inspect internal lighting for full functionality and brightness
|
Completed
|
|
|
Perform cosmetic inspection to confirm quality standards of exterior and interior
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|