IKA C-MAG HS 7 Digital Hotplate Stirrer
Temperature Range 50 to 500°C
Stirrer Speed 100 to 1,500 RPM
Platform Size 7.1 x 7.1 in (18 x 18 cm)
- Powerful motor for stirring quantities of up to 10 l (H2O)
- Simultaneous digital display of target and actual temperatures via LCD display
- Directly connection for a PT 1000 temperature sensor enable a precise temperature control (included in delivery)
- Control accuracy in medium +/- 0,5 K (in combination with PT 1000)
- Display of actual temperature value in medium with 0,1 K resolution when using PT 1000 temperature sensor
- 3 operating modes to choose (standard, safe, adjustment protection)
- Fixed safety circuit of 550°C
- Hot surface warning to prevent burns
- Digital error code display
- Elevated control panel for protection against leaking liquids
The IKA C-MAG HS 7 Digital Hotplate Stirrer is a reliable laboratory tool designed for precise and consistent heating and stirring applications. Its robust ceramic plate provides excellent chemical resistance and even heat distribution, supporting samples up to 10 liters. The digital LCD display clearly shows set and actual temperature values simultaneously, ensuring accurate control. Enhanced safety features include a fixed safety circuit at 550°C and a hot surface warning indicator. The compact, elevated control panel is protected against liquid spills, making it both safe and user-friendly for routine lab work.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The IKA C-MAG HS 7 Digital Hotplate Stirrer includes:
- > Temperature Probe
- > Support Stand
- > Operator's Manual
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
IKA C-MAG HS 7 Digital Hotplate Stirrer Specifications
|
Temperature Range
|
50 to 500°C
|
|
Stirrer Speed
|
100 to 1,500 RPM
|
|
Max. Stirring Volume
|
10L
|
|
Platform Size
|
7.1 x 7.1 in (18 x 18 cm)
|
|
Plate Material
|
Ceramic
|
|
Weight
|
11 lb. (5 kg)
|
|
Dimensions (W x D x H)
|
7.1 x 10 x 4.8 in. (21.5 x 33 x 10.5 cm)
|
|
Electrical Requirements
|
120V, 60Hz, 1020W
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
| Task |
Completed? |
Remarks? |
| Inspect display for clarity and replace if needed | Completed | |
| Examine external regulators for integrity and damage | Completed | |
| Assess temperature, CO₂, and H₂O sensors for accuracy | Completed | |
| Check internal wiring for frays, tears, or breakages | Completed | |
| Confirm software settings match internal component specs | Completed | |
| Inspect and clean fan, replacing if wear is detected | Completed | |
| Verify condition of wheels and leveling mounts | Completed | |
| Inspect door gaskets and latches for sealing integrity | Completed | |
| Calibrate temperature to 37°C with NIST-traceable probe | Completed | |
| Calibrate CO₂ to 5% with Fyrite combustion test kit | Completed | |
| Confirm 24-hour temperature and CO₂ stability | Completed | |
| Conduct cosmetic inspection to ensure appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|