Fisher Scientific Isotemp Microbiological Gravity Convection Incubator, Cat. #: 151030513
Temperature Range Ambient +5°C to 75°C
Temperature Uniformity ± 0.6°C at 37°C
- Gravity convection airflow minimizes sample drying and cross-contamination
- Microprocessor controller with large, easy-to-read fluorescent display
- Automatic overtemperature alarm enhances safety with no user settings required
- Corrosion-resistant stainless steel interior with rounded edges for easy cleaning
- Flexible shelving system allows tool-free adjustment and removal
- Internal glass door enables undisturbed sample viewing
- Silicone-sealed access port supports data logging or use of probes
- Stackable design conserves valuable bench space in multi-unit setups
The Fisher Scientific Isotemp Microbiological Gravity Convection Incubator, Cat. #: 151030513, is designed to deliver precise and stable incubation conditions for microbiological applications. With a 2.6 cu. ft. chamber capacity and a temperature range from ambient +5°C to 75°C, it provides outstanding uniformity (±0.6°C at 37°C) and stability (±0.2°C at 37°C) to maintain reproducible results. Gentle gravity convection airflow minimizes sample drying and reduces cross-contamination risks, making it ideal for applications requiring long incubation periods.
This incubator combines functionality with ease of use. Features include a corrosion-resistant stainless steel interior with rounded edges for easy cleaning, flexible shelving with tool-free adjustment, and an intuitive microprocessor controller with a large fluorescent display. Safety is reinforced with an automatic overtemperature alarm requiring no operator settings. The unit also includes an internal glass door for undisturbed sample viewing, a silicone-sealed access port for probes or data logging, and stackable construction to maximize bench space. The Fisher Scientific Isotemp Microbiological Gravity Convection Incubator, Cat. #: 151030513, is a dependable choice for accurate and efficient incubation.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Fisher Scientific Isotemp Microbiological Gravity Convection Incubator, Cat. #: 151030513 includes:
- > 2 Shelves
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Fisher Scientific Isotemp Microbiological Gravity Convection Incubator, Cat. #: 151030513 Specifications
|
Temperature Range
|
Ambient +5°C to 75°C
|
|
Temperature Uniformity
|
± 0.6°C at 37°C
|
|
Temperature Stability
|
± 0.2°C at 37°C
|
|
Volume
|
2.6 cu. ft. (75L)
|
|
Shelf Load
|
55 lb. (25 kg)
|
|
Shelf Dimensions (W x D)
|
12.95 x 14.8 in. (32.9 x 37.6 cm)
|
|
Internal Dimensions (W x D x H)
|
13.9 x 16.3 x 20.0 in. (35.4 x 41.4 x 50.8 cm)
|
|
External Dimensions (W x D x H)
|
20.9 x 22.2 x 28.3 in. (53 x 56.5 x 72 cm)
|
|
Weight
|
88 lb. (40 kg)
|
|
Electrical Requirements
|
120V, 60Hz, 2.5A, 300W
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|