Eppendorf 5424R Centrifuge w/ FA-45-24-11 Rotor & Aerosol Lid
Max Capacity 24 x 2.0ml tubes
Temperature Range -10°C to +40°C
- OptiBowl technology provides quiet operation for a comfortable laboratory environment
- Soft-touch one-finger lid closure for easy and secure access during operation
- SOFT brake mechanism ensures gentle deceleration to protect sample integrity
- Full digital display offers clear, real-time monitoring of operational parameters
- Dedicated short-spin key enables rapid centrifugation cycles for efficient processing
- Automatic lid open function enhances user convenience at the end of each run
The Eppendorf 5424R Centrifuge offers a temperature range of -10°C to +40°C, a max speed of 15,000 rpm and a max relative centrifugal force of 21,130 x g. With 24-place capacity and high max speeds, the 5424R is perfect for all modern molecular biology applications with PCR or Eppendorf tubes. The 5424R features OptiBowl for quiet operation, soft-touch one-finger lid closure, SOFT brake for samples, full digital display, separate short-spin key for quick centrifugation, and auto lid open at end of run.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Eppendorf 5424R Refrigerated Centrifuge includes:
- > FA-45-24-11 Rotor with Aerosol Lid
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Rotor Wrench
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Eppendorf 5424R Centrifuge Specifications
Maximum Speed
|
15,000 RPM
|
Maximum RCF
|
21,130 x g
|
Maximum Capacity
|
24 x 1.5/2.0ml
|
Temperature Range
|
-10°C to +40°C
|
Maximum Kinetic Energy
|
4,070 Nm
|
Timer
|
30 seconds to 9 hours 59 minutes
|
Dimensions (W x D x H)
|
11.42 x 18.90 x 10.24 in. (29 x 48 x 26 cm)
|
Weight
|
46.3 lb. (21 kg)
|
Electrical Requirements
|
120V, 50/60Hz, 3.2A, 350W
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
 |