Binder KBF 240 Humidity Test Chamber
Temperature Range 0°C to 70°C
Volume 247L (8.7 cu. ft.)
Humidity Setpoint Range 10 to 80% RH
- APT.line preheating and refrigeration system for uniform, reproducible climate conditions
- Electronically controlled humidification and dehumidification with capacitive humidity sensor
- MCS controller with 25 programmable methods and integrated electronic chart recorder
- User-friendly LCD interface with clear menu navigation and real-time clock
- Independent, adjustable temperature safety device (Class 3.1) with visual and audible alarms
- Inner glass door with sealing for stable viewing and minimal climate disturbance
- Access port with silicone plug for external probes or data-logging equipment
- Designed for long-term stability testing compliant with ICH guideline Q1A
The Binder KBF 240 Humidity Test Chamber provides highly stable, fully controlled temperature and humidity conditions for long-term stability studies and regulated testing environments. Its APT.line preheating and refrigeration system ensures uniform, reproducible climate performance, while electronically controlled humidification and dehumidification deliver precise RH regulation across a 10–80% range. Designed for reliability, the chamber maintains consistent conditions independent of the water supply, offering a complete, self-contained solution for temperature and humidity simulation.
Equipped with an intuitive MCS controller, the KBF 240 supports up to 25 programmable methods with 100 sections each, along with real-time chart recording and multiple display options for process visualization. A capacitive humidity sensor, inner glass door, independent temperature safety device (Class 3.1), and an access port for external instrumentation further enhance performance, safety, and versatility. Suitable for ICH Q1A stability testing, it serves key industries including pharmaceuticals, cosmetics, food and beverage, and packaging with dependable, long-term climate control.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Binder KBF 240 Humidity Test Chamber includes:
- > 1 Shelf
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Binder KBF 240 Humidity Test Chamber Specifications
|
Temperature Range
|
0°C to 70°C
|
|
Temperature Range (W/ Humidity)
|
10°C to 70°C
|
|
Volume
|
247L (8.7 cu. ft.)
|
|
Heat Compensation at 40°C
|
300 W
|
|
Humidity Range
|
10 to 80% RH
|
|
Temperature Variation (depending on setpoint)
|
± 0.3 K (at 25°C / 60% RH)
± 0.3 K (at 40°C / 75% RH)
|
|
Temperature Fluctuation (depending on setpoint)
|
± 0.1 K (at 25°C / 60% RH)
± 0.1 K (at 40°C / 75% RH)
|
|
Humidity Fluctuation (depending on setpoint)
|
± 1.5% (at 25°C / 60% RH)
± 1.5% (at 40°C / 75% RH)
|
|
Maximum Loading per Shelf
|
66 lb. (30 kg)
|
|
Number of Shelves (Std./Max.)
|
1 / 9
|
|
Interior Dimensions (W x D x H)
|
25.6 x 19.1 x 30.9 in. (65 x 48.5 x 78.5 cm)
|
|
Exterior Dimensions (W x D x H)
|
36.6 x 34.6 x 57.5 in. (93 x 88 x 146 cm)
|
|
Weight
|
406 lb. (184 kg)
|
|
Electrical Requirements
|
200–240V, 50/60Hz, 9.5A, 2.1 kW
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
|
Task
|
Completed?
|
Remarks?
|
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|