Beckman Coulter Allegra V-15R Refrigerated Centrifuge
Maximum Speed 13,500 RPM (4,500 RPM with included rotor)
Temperature Range -10C to 40C
Maximum Capacity 4 x 750ml
- Soft-close lid and low-profile design enhance user comfort and accessibility
- Quiet operation at ≤56 dBA supports a more peaceful lab environment
- Intuitive interface with backlit display and illuminated buttons simplifies operation
- Compact footprint saves up to 28% more bench space than comparable models
- Automatic rotor recognition and cycle tracking ensure safe, accurate runs
- Built-in imbalance detection safeguards samples and the instrument
- Rapid Temp pre-cooling function protects temperature-sensitive workflows
- ECO Mode reduces energy use and extends compressor life during standby
The Beckman Coulter Allegra V-15R Refrigerated Centrifuge is a compact, high-efficiency benchtop solution designed to handle diverse lab workflows with ease. Featuring an intuitive backlit interface with illuminated buttons, it supports up to 50 programmable runs for quick recall of common protocols. Its soft-close lid, low-profile design, and quiet operation at ≤56 dBA make it easy to integrate into busy lab environments without disruption.
Engineered for safety and reliability, the V-15R includes automatic rotor recognition, cycle tracking, and imbalance detection to protect both samples and users. The Rapid Temp pre-cooling function ensures chambers are ready when needed, while ECO Mode helps reduce energy use during idle periods. With a small footprint and powerful performance, the Allegra V-15R delivers dependable, space-saving centrifugation ideal for high-throughput and routine applications alike.
The included VS 4.750 Swing Bucket Rotor rotor reaches a maximum speed of 4,500 RPM and a maximum RCF of 4,264 x g.All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Beckman Coulter Allegra V-15R Refrigerated Centrifuge includes:
- > VS 4.750 Swing Bucket Rotor
- > Set of 4 x Hex Buckets
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Beckman Coulter Allegra V-15R Refrigerated Centrifuge Specifications
|
Maximum Speed
|
13,500 RPM (4,500 RPM with included rotor)
|
|
Minimum Speed
|
100 RPM
|
|
Maximum RCF
|
20,412 x g (4,264 x g with included rotor)
|
|
Speed Increments
|
100 RPM
|
|
Speed Accuracy
|
± 30 RPM of set speed
|
|
Maximum Capacity
|
4 x 750ml
|
|
Temperature Range
|
-10C to 40C
|
|
Timer
|
10 seconds to 99 hours 59 minutes, and 59 seconds, with continuous run (hold), and pulse modes
|
|
Acceleration Rates
|
10
|
|
Deceleration Rates
|
10
|
|
Program Memory
|
50 programs
|
|
Noise Level
|
56 dBa
|
|
Maximum Heat Dissipation
|
5527 BTU/H (1.62 kW)
|
|
Dimensions (W x D x H)
|
23.8 x 25 x 15.4 in. (60.5 x 63.5 x 39 cm)
|
|
Weight
|
243 lb. (110 kg)
|
|
Electrical Requirements
|
120V, 60Hz, 16A
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly

- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
| Task |
Completed? |
Remarks? |
| Confirm power supply output voltages match specified requirements |
Completed |
|
| Measure and verify voltages at PCB test points for accuracy |
Completed |
|
| Confirm rotor compatibility and detection for proper identification |
Completed |
|
| Inspect exterior for cosmetic or structural damage |
Completed |
|
| Check door mechanism alignment for secure opening and closing |
Completed |
|
| Inspect and replace gas strut with OEM part as needed |
Completed |
|
| Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
| Replace and confirm functionality of all lid lock switches |
Completed |
|
| Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
| Test compressor (if applicable) for proper operation and cooling |
Completed |
|
| Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
| Check rotor motor for stability, noise, and integrity |
Completed |
|
| Inspect and clear condensation drain to prevent buildup |
Completed |
|
| Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
| Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
 |
 |
 |
 |

|