Agilent 1260 Infinity I HPLC G1312B Binary Pump
Flow Range 0.05 - 5.0 mL/min
Pressure Operating Range
0 - 60 MPa (0 - 600 bar, 0 - 8,700 psi) up to 5 mL/min
Flow Precision ≤ 0.07% RSD or 0.02 min SD, whatever is greater
- Dual-piston in-series design with servo-controlled variable stroke drive
- Gear and ball screw-driven floating pistons ensure consistent performance
- High-pressure binary mixing for precise gradient formation
- Standard and low delay volume configurations available for optimized response
- Composition range settable from 0 to 100% with high accuracy and precision
- Predefined compressibility compensation enhances solvent delivery stability
- Compatible with Agilent ChemStation, EZchrom, and MassHunter software
- Recommended pH range from 1.0 to 12.5 for versatile solvent compatibility
- Integrated purge valve with PTFE frit for easy system priming
The Agilent 1260 Infinity I HPLC G1312B Binary Pump is based on a two-channel, dual-piston in-series design which comprises all essential functions that a solvent delivery system has to fulfill. Metering of solvent and delivery to the high-pressure side are performed by two pump assemblies which can generate pressure up to 600 bar.
Each channel comprises a pump assembly including pump drive, pump head, active inlet valve with replaceable cartridge, and outlet valve. The two channels are fed into a low-volume mixing chamber which is connected via a restriction capillary coil to a damping unit and a mixer. A pressure sensor monitors the pump pressure. A purge valve with integrated PTFE frit is fitted to the pump outlet for convenient priming of the pumping system. Typical applications are high throughput methods with fast gradients on high resolution 2.1 mm columns. The pump is capable of delivering flow in the range of 0.1 to 5 mL/min.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Agilent 1260 Infinity I HPLC G1312B Binary Pump includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Leak and Pressure Diagnostic Report
- > Lan Cable / Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Agilent 1260 Infinity I HPLC G1312B Binary Pump Specifications
|
Hydraulic System
|
Two dual piston in series pumps with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
|
|
Setable Flow Range
|
0.001 - 5 mL/min, in 0.001 mL/min increments
|
|
Flow Range
|
0.05 - 5.0 mL/min
|
|
Flow Precision
|
≤ 0.07% RSD or 0.02 min SD, whatever is greater
|
|
Flow Accuracy
|
± 1% or 10 µl/min, what ever is greater
|
|
Pressure Operating Range
|
0 - 60 MPa (0 - 600 bar, 0 - 8,700 psi) up to 5 mL/min
|
|
Pressure Pulsation
|
< 2% amplitude or < 3 bar at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)
|
|
Compressibility Compensation
|
Pre-defined, based on mobile phase compressibility
|
|
Recommended pH Range
|
1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
|
|
Gradient Formation
|
High-pressure binary mixing
|
|
Delay Volume
|
Standard delay volume configuration: 600 - 800 µl, (includes 400 µl mixer), dependent on back pressure
Low delay volume configuration: 120 µl
|
|
Composition Range
|
Settable range: 0 to 100%
Recommended Range: 1 to 99% or 5 µl/min per channel, whatever is greater
|
Composition Precision
|
< 0.15 % RSD or < 0.04 min SD, whatever is greater
|
|
Composition Accuracy
|
± 0.35% absolute, at 2 mL/min, at 10 MPa (100 bar)
|
|
Control
|
Agilent ChemStation, EZchrom, MassHunter
|
|
Dimensions (H x W x D)
|
7 x 13.5 x 17 in. (18 x 34.5 x 43.5 cm)
|
|
Weight
|
34 lbs. (15.5 kg)
|
|
Electrical Requirements
|
100 - 240 VAC, 50/60Hz, 74W
|
|
|
 |
 |
 |
 |

|