Agilent 1200 HPLC G1365B MWD Detector
The Agilent 1200 HPLC G1365B Multiple Wavelength Detector is a dedicated MWD that offers excellent sensitivity for
both single and multiple wavelength detection. This Multiple Wavelength Detector utilizes deuterium and
tungsten lamps with a wavelength range of 190 to 950 nm, offering an expanded
detection range with lowest detection limit. The G1365B’s diode array design enables
simultaneous acquisition of up to eight compound-specific wavelengths for
increased sensitivity and selectivity. The programmable slit (1 to 16 nm) also
allows rapid optimization of sensitivity, linearity, and spectral resolution.
All of our equipment is tested to perform at factory specifications before being shipped.
The Agilent 1200 HPLC G1365B MWD Detector includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Power Cord
- > Detector Calibration Report
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Agilent 1200 HPLC G1365B MWD Detector Specifications
|
Detection Type
|
1024-element photodiode array
|
|
Light Source
|
Deuterium and tungsten lamps
|
|
Wavelength Range
|
190 – 950 nm
|
|
Short Term Noise (ASTM) Single and Multi-Wavelength
|
± 1 × 10^-5AU at 254 and 750 nm
|
|
Drift
|
2 × 10^-3AU/hr at 254 nm
|
|
Linear Absorbance Range
|
> 2 AU (upper limit)
|
|
Wavelength Accuracy
|
± 1 nm
|
|
Wavelength Bunching
|
1 – 400 nm
|
|
Slit Width
|
1, 2, 4, 8, 16 nm
|
|
Diode Width
|
< 1 nm
|
|
Flow Cells
|
Standard: 13 µl volume, 10 mm cell path length and 120 bar (1760 psi)
pressure maximum
Semi-Micro: 5 µl volume, 6 mm cell path length and 120 bar (1760 psi)
pressure maximum
High pressure: 1.7 µl volume, 6 mm cell path length and 400 bar (5880
psi) pressure maximum
500 nano: 0.5 µl volume, 10 mm cell path length and 50 bar (725 psi) pressure
maximum
|
|
Control and Data Evaluation
|
Agilent ChemStation for LC
|
|
Analog Outputs
|
Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two
outputs
|
|
Communications
|
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready,
start, stop and shut-down signals, LAN optional
|
|
Safety and Maintenance
|
Extensive diagnostics, error detection and display (through control
module and ChemStation), leak detection, safe leak handling, leak output
signal for shutdown of pumping system. Low voltages in major maintenance areas.
|
|
GLP Features
|
Early maintenance feedback (EMF) for continuous tracking of
instrument usage in terms of lamp burn time with user-settable limits and
feedback messages. Electronic records of maintenance and errors. Verification
of wavelength accuracy with built-in holmium oxide filter
|
|
Housing
|
All materials recyclable.
|
|
|
 |
 |
 |
 |

|