Agilent 1200 HPLC G1316C Column Compartment
Temperature Range
Ambient -10°C to 100°C
Temperature Stability
± 0.05°C
Column Capacity
3 x 30 cm Columns
- Stackable, thermostatted column compartment designed for Agilent 1200 Series LC systems
- Peltier-based heating and cooling architecture for controlled column environments
- Holds up to three 30 cm columns within an optimized, low-dead-volume layout
- Two independently programmable, low-volume heat exchangers for solvent conditioning
- Heat exchanger design promotes uniform thermal conditions between solvent and column
- Electronic column identification module supports GLP documentation
- Designed for stable, reproducible chromatographic performance
- Compatible with standalone operation or integrated Agilent 1200 Series configurations
The Agilent 1200 HPLC G1316C Column Compartment is a thermostatted, stackable column compartment designed to deliver precise and stable temperature control for liquid chromatography applications where retention time reproducibility is critical. Engineered as part of the Agilent 1200 Series, the G1316C supports both heating and cooling using Peltier technology, allowing columns and mobile phase to be maintained under tightly controlled thermal conditions for consistent chromatographic performance.
The compartment accommodates up to three 30 cm columns and features an optimized internal design that minimizes dead volume while maintaining high efficiency. Temperature regulation is achieved through two independently programmable, low-volume heat exchangers that condition the solvent before it reaches the column. These heat exchangers are designed to closely match solvent and column temperatures through a combination of direct thermal contact, convection, and radiation, helping ensure uniform temperature conditions throughout the column.
Actual temperature control occurs at the heat exchanger level, with the system accounting for temperature offsets between the exchanger and column inlet. The displayed temperature reflects this corrected value, providing an accurate representation of operating conditions. An integrated electronic column identification module supports GLP documentation by recording column-specific information, while the overall thermal design allows rapid equilibration and stable operation across a wide range of flow rates and ambient conditions. Together, these features make the Agilent 1200 HPLC G1316C Column Compartment a reliable solution for applications requiring precise temperature control and reproducible chromatographic results.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Agilent 1200 HPLC G1316C Column Compartment includes:
- > Firmware Version 07.02 (Can be changed upon request)
- > Lab Advisor Diagnostic Report
- > Switching Valve
- > Power Cord/Lan Cable
- > Front Cover
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Agilent 1200 HPLC G1316C Column Compartment Specifications
|
Temperature Range
|
Ambient -10°C to 100°C
|
|
Temperature Stability
|
± 0.05°C
|
|
Temperature Accuracy
|
± 0.8°C
|
|
Column Capacity
|
Three 30 cm columns
|
|
Warm-Up / Cool-Down Time
|
5 minutes from ambient to 40°C
10 minutes from 40°C to 20°C
|
|
Dead Volume
|
3 ul heat left heat exchanger
6 ul right heat exchanger
|
|
Dimensions (W x D x H)
|
16.1 x 17 x 5.5 in. (41 x 43.5 x 14 cm)
|
|
Weight
|
22.5 lb. (10.2 kg.)
|
|
Electrical Requirements
|
100-240V, 50/60Hz, 320VA
|