Home > Chromatography >

Agilent 1100 HPLC G1365B Multi-Wavelength Detector
Agilent 1100 HPLC G1365B MWD Detector
Click Here to View the Technical Specs for the G1365B MWD
 
Alternative Views:


Our Price: $3,985.00

Condition: Fully Refurbished and Tested to Manufacturer Specifications
Warranty: 90-Day Full Parts and Labor (Click for Info)
Typically Ships Within: 3 to 7 Business Days

Request a Formal Quote


Quantity in Stock:1
Product Code: AG-MWD
Qty:

Description
Refurbishment Protocol
Technical Specification Sheet (NEW)
 

Agilent 1100 HPLC G1365B Multi-Wavelength Detector


The Agilent 1100 HPLC G1365B Multi-Wavelength Detector is a dedicated MWD that offers excellent sensitivity for both single and multiple wavelength detection. This MWD utilizes deuterium and tungsten lamps with a wavelength range of 190 to 950 nm, offering an expanded detection range with lowest detection limit. The G1365B’s diode array design enables simultaneous acquisition of up to eight compound-specific wavelengths for increased sensitivity and selectivity. The programmable slit (1 to 16 nm) also allows rapid optimization of sensitivity, linearity, and spectral resolution.


All of our equipment is tested to perform at factory specifications before being shippped.

Included with the Agilent 1100 HPLC G1365B Multi-Wavelength Detector:
Includes: > Full 90 Day Parts and Labor Warranty


Agilent 1100 HPLC G1365B Multi-Wavelength Detector Specifications

Detection Type

1024-element photodiode array

Light Source

Deuterium and tungsten lamps

Wavelength Range

190 – 950 nm

Short Term Noise (ASTM) Single and Multi-Wavelength

± 1 × 10-5AU at 254 and 750 nm

Drift

2 × 10-3AU/hr at 254 nm

Linear Absorbance Range

> 2 AU (upper limit)

Wavelength Accuracy

± 1 nm

Wavelength Bunching

1 – 400 nm

Slit Width

1, 2, 4, 8, 16 nm

Diode Width

< 1 nm

Flow Cells

Standard: 13 µl volume, 10 mm cell path length and 120 bar (1760 psi) pressure maximum

Semi-Micro: 5 µl volume, 6 mm cell path length and 120 bar (1760 psi) pressure maximum

High pressure: 1.7 µl volume, 6 mm cell path length and 400 bar (5880 psi) pressure maximum

500 nano: 0.5 µl volume, 10 mm cell path length and 50 bar (725 psi) pressure maximum

Control and Data Evaluation

Agilent ChemStation for LC

Analog Outputs

Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs

Communications

Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional

Safety and Maintenance

Extensive diagnostics, error detection and display (through control module and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.

GLP Features

Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter

Housing

All materials recyclable.



Agilent G1365B MWD Manual

Click Here to View all the Technical Specs for the G1365B MWD

Why Buy From Marshall Scientific?

Every product we offer undergoes a rigorous inspection process to verify that all manufacturer specifications are met. All of the electronics and boards are tested for proper function, and any necessary preventative maintenance is performed alongside repairs by our industry trained technicians. Our primary goal is to offer products that are as close to new as possible at an affordable price.

All Lab Equipment is Fully Refurbished and Tested

Most Pieces Ship within 2 to 5 Business Days

All Products Come with a Standard Warranty

Trained Customer and Technical Support

Industry Leading Overall Customer Satisfaction

Marshall Scientific's Certified Refurbishment Protocol

×

Our technicians decontaminate and clean all equipment upon arrival and upon purchase to create a neutral environment for lab use.

Preliminary testing is conducted to identify potential product defects, upon completion, we replace all faulty and/or under-performing parts with original manufacturer parts that perform up to specification.

Any parts known to fail or malfunction from excessive use are removed and replaced to insure the extended lifetime of your equipment.

Any necessary preventative maintenance is performed alongside repairs, so your purchase will be ready to go as soon as it arrives.

All electronics and boards are tested for proper function and response by our industry trained technicians. Any electronic deficiencies detected are immediately repaired.

Final testing cycles ensure your instrument performs fully at manufacturer specifications, guaranteeing you receive a properly working product ready for use.

Once we finish our rigorous testing and decontamination procedures, your order will be packaged to best protect it during shipment and will include delivery insurance.


Share your knowledge of this product. Be the first to write a review »

Browse for more products in the same category as this item:

Chromatography
Chromatography > HPLC
Chromatography > HPLC > Agilent
Chromatography > HPLC > Agilent > Detectors