Agilent 1260 Infinity HPLC G1311B Quat Pump Agilent 1260 Infinity HPLC G1311B Quat Pump

Agilent 1260 Infinity HPLC G1311B Quat Pump Specifications

Setable Flow Range

0.001 ƒš‚Å“ 10 mL/min, in 0.001 mL/min increments

Flow Range

0.2 ƒš‚Å“ 10.0 mL/min

Flow Precision

ƒ‚°¤ 0.07% RSD, or ƒ‚°¤ 0.02 min SD whatever is greater, based on retention time at constant room temperature

Flow Accuracy

ƒš± 1% or 10 ƒšµl/min whatever is greater

Pressure Operating Range

0 ƒš‚Å“ 60 MPa (0 ƒš‚Å“ 600 bar, 0 ƒš‚Å“ 8,700 psi) up to 5 mL/min

0 ƒš‚Å“ 20 MPa (0 ƒš‚Å“ 200 bar, 0 ƒš‚Å“ 2,950 psi) up to 10 mL/min

Pressure Pulsation

< 2% amplitude or < 3 bar at 1 mL/min isopropanol, at all pressure > 1 MPa (10 bar, 145 psi)

Recommended pH Range

1.0 ƒš‚Å“ 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel

Gradient Formation

Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve

Delay Volume600 - 900ƒšµl, dependent on back pressure

Composition Range

0 ƒš‚Å“ 95% or 5 ƒš‚Å“ 100%, user selectable

Composition Precision

< 0.2% RSD or < 0.04 min SD

Integrated Degassing UnitNumber of channels: 4

Internal volume per channel: 1.5 mL

Electrical Requirements

100 - 240V, 50/60 Hz, 180 VA

Our Price: $11,545.00
Sale Price: $8,295.00
Money Saved: $3,250.00
1 in stock!
Agilent 1260 Infinity HPLC G1311C Quat Pump VL Agilent 1260 Infinity HPLC G1311C Quat Pump VL

Agilent 1260 Infinity HPLC G1311C Quat Pump VL Specifications

Hydraulic System Dual piston in series pump with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons

Settable Flow Range

Set points 0.001 to 10 ml/min, in 0.001 ml/min increments

Flow Range

0.2 to 10.0 ml/min

Flow Precision

°¤ 0.07% RSD, or °¤ 0.02 min SD whatever is greater, based on retention time at constant room temperature

Flow Accuracy

± 1% or 10 µl/min, whatever is greater, pumping degassed H2O at 10 MPa (100 bar)

Pressure Operating Range

Operating range up to 40 MPa (400 bar, 5880 psi) up to 5 ml/min

Operating range up to 20 MPa (200 bar, 2950 psi) up to 10 ml/min

Pressure Pulsation

< 2 % amplitude (typically < 1.0%) or < 0.3 MPa (3 bar, 44 psi), whatever is greater, at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar, 145 psi)

Compressibility Compensation

User selectable, based on mobile phase compressibility

Recommended pH Range

1.0 to 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel

Gradient Formation

Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve

Delay Volume

600 to 900 µl, dependent on back pressure; measured with water at 1 ml/min

Composition Range

0 to 95% or 5 to 100%, user selectable

Composition Precision

< 0.2 % RSD or < 0.04 min SD, whatever is greater, at 1 ml/min; based on retention time at constant room temperature

Integrated Degasser

Number of channels: 4

Internal volume per channel: 1.5 ml

Our Price: $8,695.00
1 in stock!
Agilent 1200 HPLC G1311A Quat Pump Agilent 1200 HPLC G1311A Quat Pump

Agilent 1200 HPLC G1311A Quat Pump Specifications

Setable Flow Range

0.001 ƒš‚Å“ 10 ml/min, in 0.001 ml/min increments

Flow Range

0.2 ƒš‚Å“ 10.0 ml/min

Flow Precision

ƒš£ 0.07% RSD, or ƒš£ 0.02 min SD whatever is greater, based on retention time at constant room temperature

Flow Accuracy

ƒš± 1% or 10 ƒšµl/min whatever is greater

Pressure

Operating range 0 ƒš‚Å“ 40 MPa (0 ƒš‚Å“ 400 bar, 0 ƒš‚Å“ 5880 psi) up to 5 ml/min

Operating range 0 ƒš‚Å“ 20 MPa (0 ƒš‚Å“ 200 bar, 0 ƒš‚Å“ 2950 psi) up to 10 ml/min

Pressure Pulsation

< 2% amplitude (typically < 1%), at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar)

Recommended pH Range

1.0 ƒš‚Å“ 12.5, solvents with pH > 2.3 should not contain acids which attack stainless steel

Gradient Formation

Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve

Delay volume 800 ƒš‚Å“ 1100 ƒšµl, dependent on back pressure

Composition Range

0 ƒš‚Å“ 95% or 5 ƒš‚Å“ 100%, user selectable

Composition Precision

< 0.2% SD, at 0.2 and 1 ml/min

Dimensions (H x W x D)

5.5 x 13.5 x 17 in

Electrical Requirements

100-120 or 22-240 VAC, 50/60 Hz, 220 VA

Our Price: $5,995.00
1 in stock!
Agilent 1260 Infinity HPLC G1312B Binary Pump Agilent 1260 Infinity HPLC G1312B Binary Pump

Agilent 1260 Infinity HPLC G1312B Binary Pump Specifications

Hydraulic System Two dual piston in series pumps with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons
Setable Flow Range 0.001 - 5 mL/min, in 0.001 mL/min increments
Flow Range

0.05 - 5.0 mL/min

Flow Precision < 0.07% RSD or < 0.02 min SD, whatever is greater
Flow Accuracy ƒš± 1% or 10 ƒšµl/min, what ever is greater
Pressure Operating Range 0 - 60 MPa (0 - 600 bar, 0 - 8,700 psi) up to 5 mL/min
Pressure Pulsation < 2% amplitude or < 3 bar at 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 147 psi)
Compressibility Compensation Pre-defined, based on mobile phase compressibility
Recommended pH Range 1.0 - 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel
Gradient Formation High-pressure binary mixing
Delay Volume Standard delay volume configuration: 600 - 800 ƒšµl, (includes 400 ƒšµl mixer), dependent on back pressure

Low delay volume configuration: 120 ƒšµl

Our Price: $9,685.00
Sale Price: $6,295.00
Money Saved: $3,390.00
2 in stock!
Agilent 1100 HPLC G1311A Quat Pump Agilent 1100 HPLC G1311A Quat Pump

Agilent 1100 HPLC G1311A Quat Pump Specifications

Setable Flow Range

0.001 ‚ 10 ml/min, in 0.001 ml/min increments

Flow Range

0.2 ‚ 10.0 ml/min

Flow Precision

< 0.3% RSD (usually 0.15%), based on retention time, at 1 ml/min

Pressure

Operating range 0 ‚ 40 MPa (0 ‚ 400 bar, 0 ‚ 5880 psi) up to 5 ml/min

Operating range 0 ‚ 20 MPa (0 ‚ 200 bar, 0 ‚ 2950 psi) up to 10 ml/min

Pressure Pulsation

< 2% amplitude (typically < 1%), at 1 ml/min isopropanol, at all pressures > 1 MPa (10 bar)

Recommended pH Range

1.0 ‚ 12.5, solvents with pH > 2.3 should not contain acids which attack stainless steel

Gradient Formation

Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve

Delay volume 800 ‚ 1100 µl, dependent on back pressure

Composition Range

0 ‚ 95% or 5 ‚ 100%, user selectable

Composition Precision

< 0.2% SD, at 0.2 and 1 ml/min

Dimensions (H x W x D)

5.5 x 13.5 x 17 in

Electrical Requirements

100-120 or 22-240 VAC, 50/60 Hz, 220 VA

Our Price: $3,995.00
4 in stock!
Agilent 1200 HPLC G1312B SL Binary Pump Agilent 1200 HPLC G1312B SL Binary Pump

Hydraulic system

Two dual piston in series pumps with proprietary servo-controlled variable stroke drive, floating piston design and active inlet valve

Setable flow range

Setpoints 0.001 ‚ 5 ml/min, in 0.001 ml/min increments

Flow range

0.1 ‚ 5.0 ml/min

Flow precision

£ 0.07% RSD, or £ 0.02 min SD whatever is greater, based on retention time at constant room temperature

Flow accuracy

± 1% or 10 µl/min whatever is greater

Pressure

Operating range 0 400 bar (0 ‚ 5880 psi) up to 5 ml/min

Pressure pulsation

< 2 % amplitude (typically < 1 %), at 1 ml/min isopropanol, at all pressures > 1 MPa

Compressibility compensation

User-selectable, based on mobile phase compressibility

Recommended pH range

1.0 ‚ 12.5, solvents with pH < 2.3 should not contain acids which attack stainless steel

Gradient formation

High-pressure binary mixing, delay volume 180 ‚ 480 µl without mixer, 600 ‚ 900 µl with mixer, dependent on back pressure

Composition range

1 ‚ 99 % or 5 µl/min per channel, whatever is greater

Composition precision

£ 0.5% absolute

Composition accuracy

± 0.15% RSD, at 1 ml/min

Our Price: $6,495.00
Deal of the Month Price: $4,795.00
Money Saved: $1,700.00
2 in stock!
Agilent 1100 HPLC G1312A Binary Pump Agilent 1100 HPLC G1312A Binary Pump

Agilent 1100 HPLC G1312A Binary Pump Specifications

Setable Flow Range

Setpoints 0.001 to 5 ml/min, in 0.001 ml/min increments

Flow Range

0.1 to 5 ml/min

Flow Precision

°¤ 0.07% RSD or °¤ 0.02 min SD

Flow Accuracy

± 1% or 10 ul/min

Pressure

Operating range 0 to 400 bar (0 to 5880 psi) up to 5 ml/min

Pressure Pulsation

< 2% amplitude (typically < 1%) at 1 ml/min isopropanol, at all pressures > 1 MPa

Compressibility Compensation

User-selectable, based on mobile phase compressibility

Hydraulic System

Two dual pistons in series pumps with proprietary servo-controlled variable stroke drive, floating piston design and active inlet valve

Recommended pH Range

1.0 to 12.5; solvents with pH < 2.3 should not contain acides with attack stainless steel

Gradient Formation

High-pressure binary mixing, delay volume 180 to 480 ul without mixer, 600 to 900 ul with mixer, dependent upon back pressure

Composition Range

1 to 99% or 5 ul/min per channel

Composition Precision

°¤ 0.5% absolute

Composition Accuracy

°¤ 0.15% RSD at 1 ml/min